8. Troubleshooting
Symptom
Possible
Cause
Action
Unstable
Bubble is adhesion to the
Installinmeasurement solution
・
・
・
measured value
sensor
that levelisnotchanging.
Solution level is ups and
Install in solution that level is not
・
・
downs .
changing rapidly.
Solution temperature is
Install insolution that temperature is
・
・
changing rapidly.
changing rapidly.
Makesure of appropriate flow rate.
・
Loose screws
Makesure of no loose screws
・
・
Tighten screws
・
Moisture of terminals block.
Dryout terminal block
・
・
Clean terminal block
・
Abnormal
Contaminationofsensor
Cleansensor
・
・
・
・
(
)
measuredvalue
Reducing electrode/solution
refer to6.1.2
interfaceeffctive area
Install in solution that level is not
・
Sensor isfloating
changing rapidly.
・
Disconnectionofsensorcable
Makesure of no loose screws
・
・
Shortcircuitofsensorcable
Makeof the terminalsconnection.
・
・
Loose connection
・
Slow speed of
Flowrate is lower
Makesure of appropriate flow rate.
・
・
・
response
Solution concentration intank
Sufficient agitation.
Install in
・
・
(
)
is notconstant.
solution of constant concentration