5.3.5 Temperature Calibration Menu
Ω
.
Temperature measuring element of the CE-8E is Pt1000
which is high precision
So, generally it is not necessary to enter this menu.
Reference thermometer which is traceable be required to enter this menu.
When differential between measured value and reference thermometer reading is far,enter this menu.
.
Next menu
before this menu,
Caution
(
Cable Length Correction Menu must becompleted
)
Ifyouusesensorexcept10m.
(
)
from Zero Calibration Menu
Temperature Calibration Menu.
①
Push
key to enter
SEL
Temperature Calibration Menu.
Main display
shows to indicate
Push
key.
②
ENT
Main display shows temperature measuredvalue.
Wait for measuring value to become stable.
Push
keyaftermeasuredvalue become tostable.
③
ENT
The measured valueiscopied onsecondarydisplay.
Secondary display starts flashing to indicate changeable value.
Push andrelease
keys tochange the valueto reference
④
「
」
▲
▼
thermometerreading.
Onepushcountsup/downstepby step.
And keep pushing, counting up/down speeds up.
Push
key after setting desired value.
⑤
ENT
Main display flashes and shows it is defined.
Display returns to the first step of this menu.
⑥
Now, this correctionmenu iscompleted.