- 2 7 -
Rinse thoroughly with deionizedordistilled water and storedryif you donot use for a long time.
7. Error
Codes
This CE-8E is equipped with display function of each error code.
Each errorcodeisshownonSecondary Display,and Main Display flashes as follow.
DISP
ALM
MEAS CAL
1
2
1
2
7.1 Description of Error Codes
errorcode
errorcontent
ditail of error
o F
Full scale over
Measured value is over 10% of maximum of
measuringrange
u F
Full scale under
Measured value is under -10% of minimum of
measuringrange.
E-13
Abnormal
zero calibration
Offset exceeded rated value during zero calibration
E-21
Disconnection of temperature
Leadcable of temperature sensor is disconnected.
element
E-22
Short
circuit
of temperature
Leadcable of temperature sensor is short circuited.
element
E-23
Temperaturefullscale over
Temperaturemeasuredvalue is 110
or over.
℃
E-24
Temperaturefullscale under
Temperaturemeasuredvalue is
10
or less.
−
℃
E-25
Abnormal temperature calibration
Instrument cannotbecorrected error.
E-91
Systemerror
Memories of calibration andsetvalueserased.
E-92
Systemerror
Defectiveaction of AD converter.
E-93
Mistaken configration DIP SW
Set DIP. SW. No.1 OFF
No.1
When errorcodesareondisplay,transmission output is fixed hold-just before.
①
E-91 &E-92mean instrument failure. Please consult our dealers/distribuors.
②