LIQUID DETECTED

Message will appear when the liquid is
detected in sample detector test.

LIQUID NOT DETECTED

Message will appear when the liquid is
not detected in sample detector test.

LOT NO:

Allows the operator to check lot number

MENU

Menu selection.

MAINTENANCE…

Instrument is in maintenance mode.

Na/K/iCa/Cl/Li /pH

Appears when power is connected or
restored after Interruption

POSITIONING SAMPLE

Message will appear during sample
positioning.

PROBE IN SAMPLE?

Requests operator to confirm that probe is

PRESS ‘Y’ TO PROCEED

in blood.

PRESS ‘N’ TO EXIT

PLEASE INSERT PACK

Message will appear when the system has
not detected reagent pack.

PRIMING CAL A

CAL A fluid line is being primed.

PRIMING CAL B

CAL B fluid line is being primed.

PRIMING ALL

CAL A & CAL B fluid line is being primed

PRINT LAST 600 SAMPLE RESULTS

Allows operator to print 600 stored
blood results

PRINT LAST 100 QC
SAMPLE RESULTS

Allows operator to print 100 stored QC
Results

PRINT LAST 100 URINE RESULTS

Allows operator to print 100 stored
urine Results

PRINTER

Allows operator to turn printer function on or
Off.

PRINT- mV

Allows operator to select if mV is selected.
The mV readings will get flagged.

Содержание ST-200 aQua

Страница 1: ...Easy to Operate Minimal Maintenance Accurate Results Economical Auto Queue Universal Analyzer Healthcare and Diagnostic Products OPERATOR S MANUAL Welco Pump Keypad...

Страница 2: ...iCa Li pH Cl Electrolyte Analyzer OPERATOR S MANUAL SENSA CORE MEDICAL INSTRUMENTATION PVT LTD Plot No 3 Export Promotion Industrial Park Pashamylaram Sangareddy Dist Hyderabad Telangana 502307 www se...

Страница 3: ...control valve 8 b Reference housing 9 c Building the Electrodes stack 10 d Installation of tubing 11 e Reagent Pack Installation 14 C Analyzer Restoration 16 D Printer Setup 17 E Power Up 18 3 Operat...

Страница 4: ...67 D Urine 67 E Expected Values 68 5 Maintenance 69 6 Principles of Operation 70 A ISE Theory 70 7 Specifications of Na K iCa Li pH Cl 72 8 Troubleshooting 74 Introduction 74 A Flow path system 76 B F...

Страница 5: ...assembly modules 7 12 8 Pump tube installation 8 13 9 Reagent pack installation 9 15 10 Vent holes on reagent pack caps 10 15 11 Printer setup 11 17 12 Back view connection 12 18 13 Aspirating sample...

Страница 6: ...You will be guided through the display menu by pressing on keypad the Y or N button in response to questions and messages that appear on the display Calibration is automatic but can be performed on d...

Страница 7: ...perly grounded The ST 200 aQua electrolyte analyzer should not be plugged into a circuit protected by a GFI Ground Fault Interrupter The environment should be as free as possible from dust mechanical...

Страница 8: ...ST 200 aQua Operator s Manual 7 ST 200 aQua Figure 1...

Страница 9: ...lyzer components follow the procedures in this chapter New ST 200 aQua electrolyte analyzers are shipped with the fluid control valve and two pump motors a Fluid Control Valve Valve is mounted in ST 2...

Страница 10: ...comes with black lock on the seal to avoid transportation leakage of KCl Solution Remove the black lock part from the seal of Reference housing mandatory before installation Figure 3 Reference Housing...

Страница 11: ...e 5 ST 200 aQua Electrolyte analyzer comes with two blocks of 3 in 1 electrodes The block 1 comes with two types of parameter combinations i e K Na K Na iCa and the block 2 comes with 8 types of param...

Страница 12: ...tallation of tubing The ST 200 aQua Electrolyte analyzers comes with two pump motors The first pump motor is used for the flow of CAL A CAL B solutions in to the electrode housing in control with the...

Страница 13: ...ST 200 aQua Operator s Manual 12 Figure 7...

Страница 14: ...ich consists of lock 2 Rollers It consists of grove at its bottom and a knob on the front side Grove side should be inserted to the base knob 3 Pump tubing Insert the tubing as per the direction by ho...

Страница 15: ...he reagent pack in your hand and place reagent pack into back side slot of the analyzer connect the CAL A tube cap to the CAL A container and CAL B tube cap to the CAL B container respectively as show...

Страница 16: ...15 Figure 9 Note Don t place the ST 200 aQua reagent packs open to atmosphere Because the solutions gets escape from the containers Only the small vent holes on the caps should be open for air inlet a...

Страница 17: ...to maintenance mode and remove the electrode housing by disconnecting first pump outflow tube from the bubble detector flow cell pin and disconnect the second pump motor inflow red tube from the elect...

Страница 18: ...adhesives After installing the paper close the cover on the top of the housing until the door gets locked Use of the ST 200 aQua electrolyte analyzer printer requires some precautions 1 Use only the...

Страница 19: ...a grounded outlet When the ST 200 aQua electrolyte analyzer is plugged in the power is ON The Na K iCa Li pH Cl analyzer will display SENSACORE ELECTROLYTE ANALYZER Ver XXXXXX When the ST 200 aQua ele...

Страница 20: ...set calibrated frequency after four hours or eight hours as per set frequency Following is the calibration procedure A Calibration Switch ON the instrument INITIALIZING ACQUIRING RANGES Please Wait SE...

Страница 21: ...______ mV Li ______ mV pH ______ mV Cl ______ mV Displays CAL B milli volts on display E g Na 13 46 mV K 30 64 mV iCa 66 31 mV Li 67 48 mV pH 31 05 mV Cl 66 45 mV CAL A MEASUREMENT 3 cycles Air Liqui...

Страница 22: ...7 10 K 61 77 iCa 36 54 Li 76 02 pH 58 53 Cl 51 54 If any parameter is not calibrated then it will show beside that parameter in home screen So that we can clearly see the status of instrument B Analyz...

Страница 23: ...he flow cell and Press Y Please wait Positioning Sample SAMPLE MEASUREMENT Sample is positioned into the sensors Na ______ mV K ______ mV iCa ______ mV Li ______ mV pH ______ mV Cl ______ mV Displays...

Страница 24: ...ol L K 0 2 40 0 mmol L iCa 0 10 0 mmol L Cl 20 0 200 0 mmol L Li 0 5 0 mmol L pH 5 0 9 0 SERUM PLASMA CSF Na 20 0 250 0 mmol L K 0 2 40 0 mmol L iCa 0 10 0 mmol L Cl 20 0 200 0 mmol L Li 0 5 0 mmol L...

Страница 25: ...TICS 2 USER MENU 3 SETUP MENU 4 PARAMETERS 5 SELECT iCa UNIT Press 1 DIAGNOSTICS 1 DAILY CLEANER 2 MAINTENANCE 3 SENSORS 4 FLOW 5 HARDWARE 6 REAGENT PACK Press 1 DAILY CLEANER is Cleaner Ready Y Conti...

Страница 26: ...G AIR PUMPING CAL A CAL B MEASUREMENT 3 cycles of Cal B Air Liquid Air Liquid Air Liquid Na ______ mV K ______ mV iCa ______ mV Li _______ mV pH ______ mV Cl ______ mV Displays CAL B milli Volts CAL A...

Страница 27: ...AL B MEASUREMENT 3 cycles Air Liquid Air Liquid Air Liquid Na ______ mV K ______ mV iCa ______ mV Li ______ mV pH ______ mV Cl _______ mV Displays CAL B milli volts on display CAL A MEASUREMENT 3 cycl...

Страница 28: ...the ST 200 aQua in STANDBY mode reduces reagents consumption by discontinuing automatic calibrations The analyzer does however draw a small amount of calibrant A solution to prevent the electrodes fro...

Страница 29: ...ST 200 aQua Operator s Manual 28 E Sub Menus i Diagnostics...

Страница 30: ...ST 200 aQua Operator s Manual 29 ii User Menu...

Страница 31: ...ST 200 aQua Operator s Manual 30 iii Setup Menu...

Страница 32: ...ST 200 aQua Operator s Manual 31 iv Parameters Note Star Indicates Blank...

Страница 33: ...5 parts urine diluent s resulting in 1 5 dilutions Note In urine mode only Na K Cl parameters are measured but not all After dilution change Analyze type to URINE mode by following below steps Press M...

Страница 34: ...ent in Urine type as follows ANALYZE URINE Y Continue N Exit Press Y is Urine Ready Y Continue N Exit Take the diluted urine solution by the pipette drop it in the flow cell and Press Y Please Wait Po...

Страница 35: ...lows will not be reported URINE RANGE Na 35 1000 mmol L K 50 200 mmol L Cl 25 500 mmol L G Stored Results Press MENU 2 USER MENU 4 PRINT STORED RESULTS 1 Sample Results 2 QC Results 3 Urine Results Th...

Страница 36: ...has been consumed can be checked whenever required by answering Y to REAGENT PACK CONSUMPTION the percentage of the reagent pack that has been left is displayed and printed ST 200 aQua will instruct...

Страница 37: ...ou can be confident in the results obtained Run all the control levels daily Data may be stored in memory for future statistical analysis To perform quality control sample set up ANALYZE TYPE as QC in...

Страница 38: ...he Cleaning Solution by the pipette and drop it in the flow cell and Press Y Aspirating Please Wait CLEANING Performing PUMPING AIR PUMPING CAL A PUMPING AIR Clean Cycle Performed PUMPING AIR PUMPING...

Страница 39: ...splays CAL A milli Volts PUMPING AIR PUMPING CAL A PUMP CAL A TEST Pump CAL 1460 PUMPING AIR PUMPING CAL A To clean the electrodes CAL B MEASUREMENT 3 cycles Air Liquid Air Liquid Air Liquid Na ______...

Страница 40: ...libration slope instrument is ready for analysis 2 MAINTENANCE Whenever operator wants to switch off the instrument then operator should activate maintenance mode then he can switch off the instrument...

Страница 41: ...NSORS Sensors is used to check mV of the electrodes for each reagent separately Press MENU MENU 1 DIAGNOSTICS 2 USER MENU 3 SETUP MENU 4 PARAMETERS 5 SELECT iCa UNIT Press 1 DIAGNOSTICS 1 DAILY CLEANE...

Страница 42: ...SOR CAL B mV Press 2 SENSOR CAL B mV CAL B MEASUREMENT 3 cycles of Cal B Air Liquid Air Liquid Air Liquid SENSOR CAL B mV Na ______ mV K ______ mV iCa ______ mV Li ______ mV pH ______ mV Cl ______ mV...

Страница 43: ...3 VALVE TEST N Exit Press 1 REAGENT PRIMING 1 PRIME CAL A 2 PRIME CAL B 3 PRIME ALL N Exit Press 1 PRIMING CAL A Please wait Air is aspirated followed by CAL A through electrode housing PRIMING CAL A...

Страница 44: ...NG CAL B Please wait Air is aspirated followed by CAL B through electrode housing PRIMING CAL A Please wait PRIMING CAL A Priming Completed N Exit Press N Air is aspirated followed by CAL A through el...

Страница 45: ...the flow path of valve by using DI water syringe FLOW 1 REAGENT PRIMING 2 PUMP CALIBRATION 3 VALVE TEST N Exit Press 3 VALVE TEST 1 CAL A FLOW 2 CAL B FLOW 3 CAL A B In Flow 4 AIR FLOW N Exit Press 1...

Страница 46: ...A B In Flow 4 AIR FLOW N Exit Press 3 CAL A B In Flow ON 60 N STOP Clean the CAL A B flow path inlet of valve and the tubing of first pump motor before counter reaches 0 Press N CAL A B In Flow OFF N...

Страница 47: ...a UNIT Press 1 DIAGNOSTICS 1 DAILY CLEANER 2 MAINTENANCE 3 SENSORS 4 FLOW 5 HARDWARE 6 REAGENT PACK Press 5 Sample detector HARDWARE 1 SAMPLE DETECTOR 2 PUMP MODULE 3 SOFTWARE VERSION N Exit Press 1 S...

Страница 48: ...tatuses Software version HARDWARE 1 SAMPLE DETECTOR 2 PUMP MODULE 3 SOFTWARE VERSION N Exit Press 3 SOFTWARE VERSION XXXXX Date DD MM YYYY N Exit It shows the version date 6 REAGENT PACK It shows the...

Страница 49: ...PTION Press 1 REAGENT PACK ID Please Wait REAGENT PACK ID BE91132C000007E0 OEM NON OEM NON OEM DISTRIBUTOR ID XX1 LOT NO XXXXX PACK VOLUME 450 mL N Exit Press N REAGENT PACK 1 REAGENT PACK ID 2 REAGEN...

Страница 50: ...ress N REAGENT PACK 1 REAGENT PACK ID 2 REAGENT PACK INFO 3 REAGENT PACK CONSUMPTION Press 3 REAGENT PACK CONSUMPTION Please Wait REAGENT PACK CONSUMPTION Solution left 98 N Exit Consumption of reagen...

Страница 51: ...RMAL RANGES 2 CALIBRATION FREQUENCY 3 SIPPING FREQUENCY 4 PRINT STORED RESULTS 5 AUTO STANDBY 6 CORRECTION FACTOR Press 1 Please make a note that if you press the RESET RANGES then instrument reset al...

Страница 52: ...GES Press 1 SAMPLE RANGES 1 Na 2 K 3 iCa 4 Cl 5 Li 6 pH N Exit Press required parameter to be changed and enter its maximum and minimum ranges Example Press 1 SAMPLE RANGES Na MAX 150 00 Na MIN 135 00...

Страница 53: ...R Press 2 CALIBRATION FREQUENCY STATUS 4 HOURS Y 4 HOURS N 8 HOURS Press Y to select 4 hours or N to select 8 hours calibration frequency SIPPING FREQUENCY For every 30min 1 hour instrument takes 400u...

Страница 54: ...Press 1 STORED SAMPLE RESULTS Please Enter 5 Dig Patient ID XXXXX Y Continue N Clear Enter Patient ID maximum up to 5 numeric digits which you want to search Press Y STORED SAMPLE RESULTS Searching P...

Страница 55: ...Exit Depending upon the result required select the range option Press 1 RESULT 00 Y Next N Exit Press Y RESULT 01 Na 135 89 K 4 215 iCa 1 083 Cl 92 559 Li 0 00 pH 7 06 Y Print N Exit Press Y to print...

Страница 56: ...XXX Y Continue N Clear Enter Patient Id maximum up to 5 numeric digits which you want to search Press Y STORED SAMPLE RESULTS Searching P ID XXXXX Na 382 28 K 70 99 Cl 374 82 Y Print N Exit Press Y to...

Страница 57: ...TION FACTOR Press 5 for Auto Standby AUTO STANDBY STATUS ON Y ON N OFF Press Y to On and N to Off CORRECTION FACTOR USER MENU 1 NORMAL RANGES 2 CALIBRATION FREQUENCY 3 SIPPING FREQUENCY 4 PRINT STORED...

Страница 58: ...e analyzed We have got Na on higher side In this case we have to insert slope factor of 0 92 Now 0 92 will be multiplied with the result of Na e g 130 0 92 119 After adding slope factor we should run...

Страница 59: ...or 0 5 variation in K pH iCa Li In the above graph of intercept Na is reading all the three control levels higher approximately by 10 In this case we have to subtract intercept as 10 00 and verify it...

Страница 60: ...etup menu MENU 1 DIAGNOSTICS 2 USER MENU 3 SETUP MENU 4 PARAMETERS 5 SELECT iCa UNIT Press 3 SETUP MENU 1 ANALYZE TYPE 2 DATE TIME 3 PRINTER 4 PRINT PARAMETER 5 PRINT mV 6 PATIENT ID Press 1 ANALYZE T...

Страница 61: ...BLE Press 4 ANALYZE TYPE QUALITY CONTROL N Exit Press N ANALYZE TYPE 1 SERUM PLASMA 2 URINE 3 WHOLE BLOOD 4 QUALITY CONTROL 5 NON DETECTABLE Press 5 ANALYZE TYPE NON DETECTABLE N Exit Non Detectable I...

Страница 62: ...by using N button we can change the field SETUP MENU 1 ANALYZE TYPE 2 DATE TIME 3 PRINTER 4 PRINT PARAMETERS 5 PRINT mV 6 PATIENT ID Press 2 DATE TIME DATE MM DD YYYY TIME HH MM SS Y SET N NEXT Press...

Страница 63: ...IME DATE June 23 2017 TIME 11 47 Y SET N NEXT Set the Time Seconds Select Seconds 30 Y SET DATE TIME DATE June 23 2017 TIME 11 47 30 Y SET N NEXT Press N The date and time has been set PRINTER User ca...

Страница 64: ...ct the parameter by pressing appropriate number if not then press Y button to see more parameters PRINT PARAMETERS 6 pH 7 All N Exit Press corresponding number for parameters details which to be analy...

Страница 65: ...D i e press 2 for AUTO status it takes the PATIENT ID spontaneously in serial wise for each sample without giving PATIENT ID In MANUAL status i e we have to provide PATIENT ID for each sample as per r...

Страница 66: ...the parameters as per requirement arrange the electrode blocks as per parameters Note In 5th Model Li parameter depends on Na and iCa parameters If Na or iCa or both Na iCa not calibrated Li won t be...

Страница 67: ...reen topped tube must be used Do not use ammonium heparin EDTA or NaF tubes Note the time of collection 2 Mix the specimen by gently inverting and rotating the tube Do not shake 3 Analyze specimens wi...

Страница 68: ...collection 2 Mix the specimen by gently inverting the tube Do not shake 3 Centrifuge the specimen within one hour of collection Carefully remove the top plasma layer for analysis Use a Pasteur pipett...

Страница 69: ...s taking into account factors such as age sex diet and other determinants of electrolyte levels Whole Blood Serum Plasma Sodium Na 135 to 150 mmol L Potassium K 3 5 to 5 0 mmol L Ionized Calcium iCa 1...

Страница 70: ...aily cleaning solution every day mandatory All other maintenance is performed by replacing the components as described in the schedule below When you wish to perform routine or any other maintenance f...

Страница 71: ...es employ a plastic tube incorporating neutral carrier ionophores The flow through Ionized Calcium electrodes includes a plastic tube specially formulated to be selective to Calcium ions The flow thro...

Страница 72: ...E potential developed in sample solution E ISE potential developed in the calibrant solution S Electrode Slope calculated during calibration Ci x Concentration of ion I in the sample Ci s Concentratio...

Страница 73: ...rect Measurement by Ion Selective Electrode ISE Detection Range Blood Na 20 0 250 0 mmol L K 0 2 40 0 mmol L iCa 0 10 0 mmol L Cl 20 200 0 mmol L Li 0 5 0 mmol L pH 5 0 9 0 Urine Na 35 1000 mmol L K 5...

Страница 74: ...roducibility expected Day to day n 20 C V 5 C V 5 C V 5 Analysis Time 75 sec Blood 65 sec Urine Data Storage 200000 Patient results 25000 Urine results and 100 QC results Calibration Automatic or on D...

Страница 75: ...in MAINTAINANCE mode After removing the electrodes solutions pack or after troubleshooting or performing maintenance the operator must do reagent priming before calibrating the analyzer To do this pro...

Страница 76: ...200 aQua Operator s Manual 75 Troubleshooting Kit Tools to help diagnose and correct problems 10mL syringe 01 Needles 03 Test Tubes 10 Fuse 2 amp 01 Fuse 5 amp 01 Electrode seals 03 Welco pump tube 0...

Страница 77: ...d for aspirating and controlling the calibration solutions A and B from the Reagent pack by moving the lock pin as per requirement 1 While aspirating Calibrant A the valve s lock does not move because...

Страница 78: ...ment and or excessive downtime PROBLEM AREA TYPICAL ERROR MESSAGES FLOW ERRORS AIR IN CAL A AIR IN CAL B AIR IN SAMPLE AIR IN CLEANER DRIFT ELECTRODES HIGH LOW DRIFT NOISE REAGENT PACKS AIR IN CAL A A...

Страница 79: ...step Flow cell OR 2 Flow cell Enter maintenance mode and remove the electrode housing Check the flow cell if any dust particles are present in the slot clean it with daily cleaner or de ionized water...

Страница 80: ...Flow VALVE TEST Check the liquid flow through valve Flush and clear valve tubing with de ionized water by disconnecting from Reagent pack and first pump motor 4 Pump motors Pump modules are used to ge...

Страница 81: ...around by placing on solid work surface Clean the outer surface of the tubing with tissue paper Apply the grease to inner surroundings of module base softly Take a four rollers set and apply grease g...

Страница 82: ...ed to make sure they are dry and clean including the reference electrode pin and female connector if low or erratic readings occur Electrodes with low calibration values below 35 for Na K Li Cl and be...

Страница 83: ...t the electrode blocks pins including the reference electrode pin to ensure that they are dry and clean After reassembly the analyzer should be primed and calibrated D Reagent Pack AIR IN CAL A AIR IN...

Страница 84: ...1 and 2 are successful but the printer remains inactive remove the analyzer power cord for ten seconds and then reinstall the power cord Press the paper advance button The printer head should move acr...

Страница 85: ...utively on the ST 200 aQua and record the data 3 Calculate the mean standard deviation and C V for each sample 4 Compare the C V for sodium potassium and chloride and the S D to specifications CORRELA...

Страница 86: ...Insufficient sample AIR IN CAL A Cal A not detected correctly AIR IN CAL B Cal B not detected correctly ASPIRATING Sample is being drawn into probe AUTO STANDBY Auto standby on off CALIBRATION FREQUEN...

Страница 87: ...to perform daily cleaning DIAGNOSTICS Allows operator to check key analyzer Functions DISTRIBUTOR ID Allows the operator to check distributor ID DRIFT Na K iCa Cl Li pH Na K iCa Cl Li pH mV drifting...

Страница 88: ...ests operator to confirm that probe is PRESS Y TO PROCEED in blood PRESS N TO EXIT PLEASE INSERT PACK Message will appear when the system has not detected reagent pack PRIMING CAL A CAL A fluid line i...

Страница 89: ...essage will appear when liquid is pumped PUMP SPEED FAST Allows the operator to set the pumping speed fast PUMP SPEED SLOW Allows the operator to set the pumping speed slow REAGENT CONSUMPTION Allows...

Страница 90: ...s the operator to display the mV readings of CAL A reagent selected SENSOR CAL B mV Allows the operator to display the mV readings of CAL B reagent selected SYSTEM UNCALIBRATED This allows the operato...

Страница 91: ...y applicable for ST 200 aQua under normal operating conditions and maintained in accordance with the instructions in the operator s manual and when not having been subjected to accident alteration or...

Страница 92: ...Auto Queue Universal Analyzer ST 200 a ua Plot No 3 Export Promotion Industrial Park Pashamylaram Sangareddy Dist Hyderabad 502307 Telangana INDIA Telephone 08455 223400 web www sensacore com...

Отзывы: