ST-200 aQua
Operator’s Manual
83
Note:
Make sure that the flow cell cover should be kept on the top of the
flow cell/bubble detector if the analyzer is kept in ideal state to protect
from the dust.
Figure 20
Figure 21
F. Printer
If the ST-200 aQua printer fails to print, attempt the following
procedures:
1. To verify that the printer is turned on, enter the SETUP MENU and
check PRINTER ON/OFF.
2. Check the printer mechanism for any paper obstructions in the
printer head area. Remove any obstructions if possible, without
damaging the printer assembly or paper tear bar.
3. If steps 1 and 2 are successful, but the printer remains inactive,
remove the analyzer power cord for ten seconds, and then reinstall
the power cord. Press the paper advance button. The printer head
should move across the paper and the paper should advance
forward.
If the printer continues to function abnormally, contact your
ST-200 aQua distributor.
Performance Verification
Periodically, the laboratory may want to verify the analyzer’s performance.
ST-200 recommends the following procedures:
CALIBRATION – Automated
1 Press “Y” to SYSTEM NOT CALIBRATED PLEASE CALIBRATE.
2 Calibrant B and Calibrant A from the solutions pack are aspirated
and measured internally.
3 ANALYZE SAMPLE indicates a successful calibration. If an error is
displayed, refer to instructions in the troubleshooting section.
Содержание ST-200 aQua
Страница 8: ...ST 200 aQua Operator s Manual 7 ST 200 aQua Figure 1...
Страница 13: ...ST 200 aQua Operator s Manual 12 Figure 7...
Страница 29: ...ST 200 aQua Operator s Manual 28 E Sub Menus i Diagnostics...
Страница 30: ...ST 200 aQua Operator s Manual 29 ii User Menu...
Страница 31: ...ST 200 aQua Operator s Manual 30 iii Setup Menu...
Страница 32: ...ST 200 aQua Operator s Manual 31 iv Parameters Note Star Indicates Blank...