Loading the Stack
Slide the electrode stack into the electrode housing. Push down firmly
until the bubble detector snaps into position.
Carefully install the bubble detector and the electrode blocks into the
electrode housing in such a manner that electrode blocks silver pins
should be inserted in female connector.
Do not force the electrode housing into place. If the electrode blocks are
properly installed, the electrode housing should slide easily into position.
d. Installation of tubing:
The ST-200 aQua Electrolyte analyzers comes with two pump motors.
The first pump motor is used for the flow of CAL-A & CAL-B solutions in to
the electrode housing in control with the valve. Its inflow is from the valve
and outflow is to the bubble detector flow cell and electrode housing.
Connect its outflow tube at the side of the flow cell pin.
The second pump motor is used for the flow of sample, QC, CAL-A & CAL-
B fluids from electrode housing to the waste bottle. Its inflow is from
electrode housing and outflow is to the waste bottle. Its pump tube has
two stoppers. One stopper (red) is connected at the base of the electrode
housing and the other stopper (blue) is connected to the waste bottle.
Содержание ST-200 aQua
Страница 8: ...ST 200 aQua Operator s Manual 7 ST 200 aQua Figure 1...
Страница 13: ...ST 200 aQua Operator s Manual 12 Figure 7...
Страница 29: ...ST 200 aQua Operator s Manual 28 E Sub Menus i Diagnostics...
Страница 30: ...ST 200 aQua Operator s Manual 29 ii User Menu...
Страница 31: ...ST 200 aQua Operator s Manual 30 iii Setup Menu...
Страница 32: ...ST 200 aQua Operator s Manual 31 iv Parameters Note Star Indicates Blank...