C A K LOW, DRIFT message may indicate that the block 1 needs replacing.
Before doing so, perform a daily cleaning cycle, as described under
DAILY CLEANER. Next, perform a calibration. If the potassium
calibration value remains low, replace block.
D A LOW, DRIFT message for chloride or Ionized Calcium only, may
indicate that replacement of the particular block is necessary. Before
doing so, try performing a daily Cleaning, as described under DAILY
CLEANER. Next, perform a calibration. If the calibration value remains
low, replace the particular electrode block.
E If all parameters have low or high calibration values, the problem is
most likely with the electrode assembly or the reference electrode (on
the Na/K/iCa/Cl/Li/pH analyzer, the problem may be the disposable
reference electrode). It is also possible that an incorrect reagent pack
is being used, the ST-200 aQua will not calibrate successfully.
F
If drift messages appear for all parameters, there is most likely a flow
problem (refer to flow problems) or electrode stack assembly problem
(refer to electrode stack assembly). Also, inspect the electrode blocks
pins, including the reference electrode pin, to ensure that they are dry
and clean. After reassembly, the analyzer should be primed and
calibrated.
D. Reagent Pack
AIR IN CAL A, AIR IN CAL B, REAGENT PACK LOW, INSUFFICIENT PACK
PLEASE REPLACE & EMPTY PACK PLEASE REPLACE messages may
indicate that the reagent pack is empty or low.
The ST-200 aQua has an RFID (Radio Frequency Identification System)
which keeps track of the reagent pack usage. The ST-200 aQua will
indicate the availability of reagents in a decreasing manner i.e., 100% to
0% at every analysis.
When 10-15% of the reagent pack has remained in the reagent pack,
instrument will indicate INSUFFICIENT PACK PLEASE REPLACE till it
reaches 5%. Below 5% the instrument will displays EMPTY PACK PLEASE
REPLACE.
E. Bubble Detector
The bubble detector detects the difference between liquid and air, which is
important for proper sample positioning in the ST-200 aQua. It consists
the flow cell at the top to drop the sample by the pipette. To verify
performance, proceed to *DIAGNOSTICS* followed by sub menu
HARDWARE and perform SAMPLE DETECTOR test.
In sample detector test, if liquid or air is not detected, clean the Bubble
detector with diluted Sodium Hypochlorite (NaClO) or Daily cleaner or de-
ionized water. Then also if not solved, replace bubble detector.
Содержание ST-200 aQua
Страница 8: ...ST 200 aQua Operator s Manual 7 ST 200 aQua Figure 1...
Страница 13: ...ST 200 aQua Operator s Manual 12 Figure 7...
Страница 29: ...ST 200 aQua Operator s Manual 28 E Sub Menus i Diagnostics...
Страница 30: ...ST 200 aQua Operator s Manual 29 ii User Menu...
Страница 31: ...ST 200 aQua Operator s Manual 30 iii Setup Menu...
Страница 32: ...ST 200 aQua Operator s Manual 31 iv Parameters Note Star Indicates Blank...