EM 100 G2, EM 300 G2, EM 500 G2
08/2004
5.3
Test table
Sequence Type of measure-
ment/adjustment
Signal input
Settings
Measuring
point/test
point
Required
value
Remarks
1
Visual inspection
1.1
Jack socket PHONES
(only EM 500 G2)
–
–
–
–
Contacts are
not twisted, no
hairline cracks
at the solder
points, no
mechanical
damage
1.2
Volume control
(only EM 500 G2)
–
–
–
–
See 1.1
1.3
왖왔
, SET,
POWER/ON buttons
–
–
–
–
See 1.1
1.4
DC socket, DC IN
–
–
–
–
See 1.1
1.5
XLR socket
–
–
–
–
See 1.1
1.6
Jack socket
–
–
–
–
See 1.1
1.7
RJ45 DATA socket
–
–
–
–
See 1.1
1.8
BNC sockets
–
–
–
–
See 1.1
1.9
Monitoring of all
display elements
+12V
DC
at DC socket
"POWER" = ON
–
–
Monitoring of
all display
segments
1.10
Monitoring of the il-
luminated
왖왔
, SET,
ON buttons
(not EM 100 G2)
See 1.9
See 1.9
–
–
–
2
Mainboard: current and voltage measurement
2.1
Standby
See 1.9
"POWER" = Off
"VOL" = min. (R1)
"AF level" = max.
–
50mA
Set laboratory
power supply
unit current
limiter to
I
ON
= 1A
2.2
Current consumpti-
on during operation
See 1.9
"POWER" = On
"VOL" = min. (R1)
"AF level" = max.
–
200mA
See 2.1
2.3
Commissioning
See 1.9
VOL = min. (R1)
AF LEVEL = max.
–
–
–
2.4
Operating voltage
(+12V
DC
)
See 1.9
POWER = On
Mainboard,
solder side:
TP209 (+)
TP208 (-)
+12V
DC
–
2.5
Operating voltage
(+6V
DD
)
See 1.9
See 2.4
TP226/ GND
+5.8V to
+6.2V
–
2.6
Operating voltage
(+5V
CC
)
See 1.9
See 2.4
TP231/ GND
+4.8V to
+5.2V
–
2.7
Operating voltage
(+5V
SW
)
See 1.9
See 2.4
TP233/ GND
+4.8V to
+5.2V
–
2.8
Operating voltage
(+6V
SW
)
See 1.9
See 2.4
TP229/ GND
+5.8V to
+6.2V
–