Note:
-
When the software is opened, the fields are grayed out and before it can be used, the “OPEN”
button must be clicked.
-
Without clicking the “Write” button, configured and customized settings will not be applied and
neither be saved.
-
If the USB connection is successful, the “Success” icon appears. If the connection fails, reconnect
the USB cable or check the device manager to see the connection status.
5.1. Software Overview
5. Software Manager
Read
Write
Calibration
Log
Firmware
•
Sensor Type
– The current sensor type in the device (CO
₂
, cannot be changed)
•
Serial Number
– iGas Detector CO2 serial number
•
Firmware Version
– Current firmware version of the unit (can change by upgrading)
•
User ID/Message
- The User ID can be used to add a usage message
•
Unit
– Adjust measuring unit by PPM or %vol
•
Stealth
– Disable the alarm, buzzer, and LED for a special occasion
•
Calibration Use
– Disable / Enable the calibration access for a special occasion
•
Auto Calibration use
– Auto “Fresh Calibration” is activated every 3 days.
•
Low Alarm & High Alarm
– The 1
st
and 2
nd
alarm set points (Min/Max: 400ppm
(0.04%vol) ~ 49,999ppm (5%vol)
•
STEL Alarm & TWA Alarm
– Short Term Exposure Limit and Time Weighted
Average level of concentration of CO
₂
(Min/Max: 400ppm (0.04%vol) ~ 49,999ppm
(5%vol)
•
Gas Concentration
– This allows a user to enter/amend correct concentration of the
gas cylinder (Min/Max: 400ppm (0.04%vol) ~ 49,999ppm (5%vol)
•
Calibration Interval (day)
– The calibration reminder informs every fixed day (can
adjust 0 (n/a) ~ 365)
•
Bump Interval (Days)
– The Bump test reminder informs every fixed day (can adjust
0 (n/a) ~ 365)
•
Self Interval (Days)
– The Self test reminder informs every fixed day (can adjust 0
(n/a) ~ 365) *Default is N/A