1. Move to a fresh air environment, which is free of hazardous gas
2. Press and hold down the power button for approximately 2 seconds until the gas type
(CO2) is displayed.
3. Upon activation, gas type(CO2), firmware version, and display appears, and the
detector performs the self diagnostic test.
4. After self test is successful, the detector countdown is displayed for 10 seconds.
5. The detector displays current CO2 concentration.
If the stabilization is failed, Error mark is displayed and measuring mode will not be entered. In this case,
contact authorized resellers or SENKO service center at 82-31-492-0445 for repair or return.
Err-1
Initial setting error
Err-2
Sensor error
Err-3
Memory error
L-bat
Low battery
Err marks
OFF
2.1. Bump test
1. Before daily use, users are required to perform bump test to see a sensor responds to
a CO2 gas.
2. To perform the bump test, follow the below steps.
- Prepare a CO2 gas over low and high alarm.
- Press and hold the pushbutton and power button for three seconds in the measurement
mode. Press the pushbutton until “TEST” is displayed and press the power button 2sec
to enter the mode.
-
Press the pushbutton until “BTS is displayed and press the power button to activate it.
- After pressing the power button, apply a CO2 gas over low and first alarm and the 30
seconds count down is displayed.
-
Once the test is passed, “SUC”(V) icon appears on the display. If test is failed, “FA”*(X)
mark appears on the display.