background image

1. Move to a fresh air environment, which is free of hazardous gas
2. Press and hold down the power button for approximately 2 seconds until the gas type 

(CO2) is displayed. 

3. Upon activation, gas type(CO2), firmware version, and display appears, and the 

detector performs the self diagnostic test.  

4. After self test is successful, the detector countdown is displayed for 10 seconds. 
5. The detector displays current  CO2 concentration. 

If the stabilization is failed, Error mark is displayed and measuring mode will not be entered. In this case, 

contact authorized resellers or SENKO service center at 82-31-492-0445 for repair or return. 

Err-1

Initial setting error 

Err-2

Sensor error

Err-3

Memory error

L-bat

Low battery

Err marks

OFF

2.1. Bump test

1. Before daily use, users are required to perform bump test to see a sensor responds to 

a CO2 gas. 

2. To perform the bump test, follow the below steps.
- Prepare a CO2 gas over low and high alarm. 
- Press and hold the pushbutton and power button for three seconds in the measurement 

mode. Press the pushbutton until “TEST” is displayed and press the power button 2sec
to enter the mode. 

-

Press the pushbutton until “BTS is displayed and press the power button to activate it.

- After pressing the power button, apply a CO2 gas over low and first alarm and the 30 

seconds count down is displayed. 

-

Once the test is passed, “SUC”(V) icon appears on the display. If test is failed, “FA”*(X) 
mark appears on the display. 

2. Activation & Deactivation

5

Содержание iGas Detector CO2

Страница 1: ...Portable Single Gas Detector User s Manual...

Страница 2: ...sensor regularly by the gas concentration exceeding alarm set points Test LED audio and vibration manually If the temperature changes sharply during use of the device e g indoors vs outdoors the valu...

Страница 3: ...etting Mode 4 1 Alarm Adjust alarm setpoints 4 2 Calibration 4 3 Clr max 5 Software Manager 6 Maintenance 4 5 6 8 9 10 11 13 16 7 Specification 18 8 Battery Charging 18 3 3 Menu Tree 7 4 4 Clr STEL an...

Страница 4: ...ration Remaining Battery 1st Alarm Display 2nd Alarm Display Alarm Condition Max Value Short Term Exposure Limit Time Weighted Average Test Success Test Fail Over Limit End of Test Part Description LC...

Страница 5: ...r 1 Initial setting error Err 2 Sensor error Err 3 Memory error L bat Low battery Err marks OFF 2 1 Bump test 1 Before daily use users are required to perform bump test to see a sensor responds to a C...

Страница 6: ...oint STEL alarm setpoint TWA alarm setpoint Firmware version Calibration concentration To move to the next menu press the pushbutton At the last step press pushbutton or do not push any button for 10...

Страница 7: ...Factory Default Test 2 sec Fail Click Click Measuring mode Click Next mode 2 sec Setting process Cal Setting Process Value Clear Process Value Clear Process Setting process Setup Testing progress 2 se...

Страница 8: ...ncentration Unit vol ppm Concentration unit conversion Init Reset Test Self Bts Self test Bump test 8 In the setting mode users can adjust setpoints perform calibration and reset previous values 1 To...

Страница 9: ...bration per second Adjust alarm setpoints To enter the setting mode press and hold the pushbutton and power button simultaneously for two seconds In the alarm setting icon press and hold down the powe...

Страница 10: ...e CO2 gas N2 or fresh air calibration are also available for calibration Check with your safety managers to ensure calibration accessibility Note that Fresh air calibration should be performed out of...

Страница 11: ...peak concentration in the detector follow below steps 1 Press the pushbutton power button simultaneously and the pushbutton until Clr max is displayed 2 Press the power button to clear the peak value...

Страница 12: ...Press the pushbutton until Test is displayed 2 Press the power button for three seconds In the SELF display press the power button for three seconds to activate the self test While it s activated the...

Страница 13: ...e message Unit Adjust measuring unit by PPM or vol Stealth Disable the alarm buzzer and LED for a special occasion Calibration Use Disable Enable the calibration access for a special occasion Auto Cal...

Страница 14: ...e log files will 5 1 2 Write 5 1 3 Calibration 5 1 4 Log 1 Connect the unit to the PC using instrument s USB 2 Plug the calibration cap not for Fresh air calibration and open the software 3 Click Cali...

Страница 15: ...e unit Test order LED Beep Motor Flash Sensor End Factory Default F Reset original settings and specifications Time Write T To set a time by user location see 8 2 3 1 Power OFF P Turn off the device I...

Страница 16: ...safety function Replacing the sensor and battery should be performed by authorized sellers agents distributors or safety managers The sensors published by SENKO should be used for replacement Product...

Страница 17: ...bration with CO2 2 vol Before use stabilize the detector for 5 minutes 6 1 Sensor Replacement 6 2 Battery Replacement 1 Move to a fresh air environment and power off the detector 2 Remove the back cas...

Страница 18: ...0g Operating Life 14 days Expected in measuring mode without alarm Event Log Recent 30 alarms Approval EMC directive 2014 30 EU Gas Measuring range Low Alarm High Alarm Resolution CO2 0 5 0 vol 0 5000...

Страница 19: ...nufacturer is not liable under this warranty if its testing and examination disclose that the alleged defect in the product does not exist or was caused by the purchaser s or any third party s misuse...

Отзывы: