In Manual calibration mode, proceed as follows for calibration:
1)
Set the Tare value in technical units
2)
Press the "SET MANUAL TARE (FLASH)" button
ATTENTION!
To obtain better measurement accuracy, use the tare acquisition mode from the field.
CELL CALIBRATION WITH A SAMPLE WEIGHT
In cell calibration with a sample weight it is necessary to have or know:
-The cell sensitivity
-The cell full scale
- A sample weight (such that its gross weight is as close as possible to the full scale)