10.2.13.3.
STABLE WEIGHING
In this mode the output is activated when the stable weighing condition is verified.
10.2.13.4.
MODBUS CONTROLLABLE
In this mode the output can be controlled directly by modbus (see STATUS register in the chapter 16.6).
10.2.13.5.
EXCEEDED THRESHOLD WITH HYSTERESIS (ALARM)
In this mode the output is activated when the net weight reaches the threshold, the alarm is cancelled when
the net weight falls below the Threshold-Hysteresis value:
STABLE WEIGHING CONDITION
The stable weighing condition is used to indicate that the net weight measurement is stable if:
The net weight remains within the weight
∆𝑝𝑒𝑠𝑜_𝑛𝑒𝑡𝑡𝑜
over time
∆𝑡𝑒𝑚𝑝𝑜
or if the slope of the curve drawn
by the net weight is less than
∆𝑝𝑒𝑠𝑜_𝑛𝑒𝑡𝑡𝑜
∆𝑡𝑒𝑚𝑝𝑜
: