17
the
recommended
separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic
site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Note
1. At 80 MHz and 900 MHz, the higher frequency range applies.
2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the ACCUNIQ BC380 is used
exceeds the applicable RF compliance level above, the ACCUNIQ BC380
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or
relocating the ACCUNIQ BC380.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.