25
Troubleshooting
This chapter presents operating problems that may occur while the device is in use.
You can check out technical tips that may help solve the problem, if it is possible (Table 3). Or, you can
contact our AUTHORIZED TECHNICAL ASSISTANCE.
PROBLEM
POSSIBLE SOLUTION
The device will NOT turn
on
• Make sure the AC power cable is properly connected
• Check the fuses
• Make sure there is current in the power outlet
The device stops working
•
If the device is switching off randomly while in operation, first check for damage in
the power cable, plug or outlet. It might be necessary to install a surge protector or,
preferably, an insulated transformer.
• If the device continues switching off randomly, call Seiler Instrument.
• Warning: Never disconnect the power cable while the device is operating. This may
cause the internal components to fail or burn.
• Warning: Never pull on the power cable to disconnect
The device works
chaotically
• Make sure there is current in the local installation of the device. This can be
done by measuring the network voltage. It must NOT fluctuate by more than 20%
compared to the specified nominal voltage (90-240 V~). Try to adjust your clinic’s
supply voltage.
• If the device continues failing intermittently, check whether some other device
installed next to it is emitting an electromagnetic field above that specified in a
standard.
• If these solutions fail, please contact our AUTHORIZED TECHNICAL
ASSISTANCE.
Table 3: Most common problems that might occur during use of the Seiler Evolution Zoom Microscope.