EcoStar Info, EcoStar Auto
Trouble-shooting
21
Message
displayed
Description
Solution
Er 7
Turbine problem.
Unplug the device from all sources of electrical power.
Check that there is no foreign object in the air outlet. Plug
in the machine again and turn it on.
If the error persists, return the machine to the technical
support.
Er 8
Voltage problem.
Unplug the device from all sources of electrical power.
Plug it in again and turn it on.
If the error persists, return the machine to the technical
support.
Er 10
Machine parameter
problem.
Unplug the device from all sources of electrical power.
Plug it in again and turn it on.
If the error persists, return the machine to the technical
support.
Er 11
High current
consumption.
Unplug the device from all sources of electrical power.
Plug it in again and turn it on.
If the error persists, return the machine to the technical
support.
Er 12
Device memory is
empty.
Unplug the device from all sources of electrical power.
Plug it in again.
If the error persists, return the machine to the technical
support.
Er 13
I²C bus communication
error.
Unplug the device from all sources of electrical power.
Plug it in again and turn it on.
If the error persists, return the machine to the technical
support
Note:
When an error is detected, the device goes into standby mode and access to the different menus is impossible.