EcoStar Info, EcoStar Auto
Setting EcoStar Auto device
15
Press
This screen appears:
Pressure decrease speed
,
available in APAP mode
only.
You can set the speed - slow (S) or fast (F) - by
pressing the button
or
.
Press
This screen appears:
Maximum pressure on apnea command
, available
in APAP mode only.
Increase or decrease the value displayed pressing
the button
or
.
Press
This screen appears:
Hour counter
(device operating time). The value
displayed on the right below represents the
minutes.
Press
This screen appears:
Compliance counter
(time during which the patient
actually breathed with the mask). The value
displayed on the right below represents the
minutes.
Press
This screen appears:
Built-in
pressure sensor offset.
Increase or decrease the value displayed using the
button
or
.
Refer to "Adjustment of the pressure at the device
outlet" page 17.
Press
This screen appears:
Selection of
RS232 connection for an external
module
(for example a modem)
The connection can be enabled (1) or disabled (0) by
pressing the button
or
.
Press
This screen appears:
Software version
included with the EcoStar Auto
device. The version can be viewed but not modified.