20
Trouble-shooting
EcoStar Info, EcoStar Auto
Trouble-shooting
Information messages
Message
displayed
Possible cause
Solution
In 1
The mask is
disconnected.
Check the connections between the mask, the tubing,
and the device. This message disappears when you
breathe in your mask well reconnected or when one of
the three buttons of the device is pressed. Otherwise, the
device will stop after 30 minutes.
Error messages
Message
displayed
Description
Solution
Er 1
Patient settings
problem.
Reinitialize the default settings by holding down the
information access button
for 2 seconds, the
blinking "DEF" message will appear. Accept by pressing
the On / Off button.
. The device is now set to the
default values and can be reset to the desired values.
If the error persists, return the machine to the technical
support.
Er 2
High motor
temperature.
Unplug the device from all sources of electrical power.
Allow it to cool and plug it in again according to the
instructions in the patient manual.
If the error persists, return the machine to the technical
support.
Er 3
Program battery
overflow.
Unplug the machine from all electrical power. Plug it in
again.
If the error recurs, return the machine to the technical
support.
Er 4
High pressure.
Unplug the machine from all electrical power. Plug it in
again.
If the error recurs, return the machine to the technical
support.
Er 6
CheckSum code error.
Unplug the device from all sources of electrical power.
Plug it in again and turn it on.
If the error persists, return the machine to the technical
support.