8
Accessing settings
EcoStar Info, EcoStar Auto
Setting ranges
The following tables indicate the parameters
adjustable and the features available on the
EcoStar Info and EcoStar Auto devices, as well as
the pneumatic calibration regarding the
accessories used.
Notes:
As the device is equipped with a differential
pressure sensor, the pressure compensation
for altitude is automatic.
The default pressure unit is cmH
2
O.
Operating mode on the EcoStar Auto:
Parameter
Display
Default
Operating mode
CPAP
APAP
(Auto-CPAP)
APAP
(Auto-CPAP)
Parameters adjustable on the devices:
Parameter
Minimum
value
Maximum
value
Default
value
Increment
On EcoStar Info and EcoStar Auto devices in CPAP mode:
Pressure level
4 cmH
2
O
20 cmH
2
O
8 cmH
2
O
0.5 cmH
2
O
Comfort pressure
4 cmH
2
O
prescribed
pressure
4 cmH
2
O
0.5 cmH
2
O
Ramp time
0 min
45 min
15 min
5 min
On EcoStar Auto device in Auto-CPAP mode:
Maximum pressure
4 cmH
2
O or
minimum
pressure
20 cmH
2
O
20 cmH
2
O
0.5 cmH
2
O
Minimum pressure
4 cmH
2
O
20 cmH
2
O or
maximum
pressure
4 cmH
2
O
0.5 cmH
2
O
Comfort pressure
4 cmH
2
O
20 cmH
2
O or
maximum
pressure
4 cmH
2
O
0.5 cmH
2
O
Latency time
0 min
45 min
15 min
5 min
Maximum pressure on apnea
command
minimum
pressure
maximum
pressure
10 cmH
2
O
0.5 cmH
2
O
Parameter adjustable on EcoStar Auto device:
Parameter
Display
Default
Pressure decrease speed (in Auto-CPAP
mode
only
) :
Sd:F
(fast)
Sd:S
(slow)
Sd:S
(slow)