Enclosure Leakage Test
26
Class I
0.5 mA
0.5 mA
0.5 mA
Class II
0.5 mA
0.5 mA
0.5 mA
Enclosure Leakage, single fault, earth open
This test measures the enclosure leakage current with a single fault condition (earth open). The current is
measured through the Measuring Device with S1 closed, S8 open and S5 in normal and then S5 reversed. The
maximum allowed values are as follows:
Type B
Type BF
Type CF
Class I
0.5 mA
0.5 mA
0.5 mA
Class II
n/a
n/a
n/a