17
After this has been done, all test results are then printed until the rotary switch is placed in the
START /
FINISH
position again and the START button is pressed to initiate the final part of the results print-out.
The user is required to manually write such items as the appliance number, date etc, and to sign off the printout
where necessary. It is also necessary for the user to manually determine whether the EUT has passed all tests.
Equipment Type
Rotating the switch to the equipment class and type symbols enables the user to select the class and type of
the equipment to be tested. The display will indicate the current equipment type setting according to the
following list:
“I : b” – Class 1 Equipment with Type B Applied Parts.
“I :bF” – Class 1 Equipment with Type BF Applied Parts.
”I :cF” – Class 1 Equipment with Type CF Applied Parts.
“II: b” – Class 2 Equipment with Type B Applied Parts.
“II:bF” – Class 2 Equipment with Type BF Applied Parts
”II:cF” – Class 2 Equipment with Type CF Applied Parts
By pressing the up/down arrows the user can cycle between different equipment class and types that will flash
on the display. Pressing the SELECT key will cause the equipment type to stop flashing indicating that the
equipment class and type had been selected.
During tests where the equipment class and type is displayed the operator has the opportunity of changing the
equipment class and type selection by pressing the SELECT button. After pressing SELECT the arrow keys are
used to cycle between the equipment class and types. Pressing SELECT again selects the current flashing
option.
Other user settings can be set following selection of the equipment type. Following equipment selection, the
user is also given the option to indicate the pass/fail limit or test sequence number when a leakage fails a given
test limit (“PASS on”/”PASS off”).
The user is given the option to enable or disable DC-Only tests for the Patient Leakage and the Patient Auxiliary
Current tests during semi-automatic tests. The DC-Only is enabled from power up.
The user is given the option to enable the Manual Tests [SFC] for the leakage tests which includes Earth
Leakage, Enclosure Leakage, Patient Leakage, Patient Leakage - Applied Part to Earth, Patient Leakage - F-
Type, Patient Leakage - Mains on SIP/SOP and Patient Auxiliary Current.
Note: These settings are NOT stored in memory on powering down the Analyser.
Earth Continuity Tests
Switch Position
------------
Earth Bond
Test Icon Displayed ---