Manual revision 005
Appendix II: Installation/Removal from Optional Sensor Mount
SBE 63
42
Appendix II: Installation/Removal from
Optional Sensor Mount
Installation
1.
Align the 4-pin connectors and push the SBE 63 onto the mount.
2.
Install the two #10-24, 1/2” titanium cap screws from the back of the
sensor mount to secure the SBE 63 in the mount.
Removal
1.
Remove the two #10-24, 1/2” titanium cap screws from the back of the
sensor mount that secure the SBE 63 in the mount.
2.
Pull the SBE 63 straight out of the sensor mount. If it is difficult to
remove:
•
Use a rubber glove to provide a better grip as you pull.
OR
•
Insert
one
of the cap screws in the sensor mount, and thread it just
one or two turns. Push the head of the screw with a driver; this will
push against the housing and free the O-ring seal. Remove the cap
screws, and then pull the SBE 63 straight out of the sensor mount.
SBE 63 – 4-pin
connector
Sensor mount –
4-pin connector
Install
cap
screws
(2)