Manual revision 005
Section 4: Deploying and Operating SBE 63
SBE 63
30
3.
Install the I/O cable on the SBE 63 sensor mount:
A.
Lightly lubricate the inside of the cable connector with silicone
grease (DC-4 or equivalent).
B.
Standard Connector
- Install the cable connector, aligning the raised
bump on the side of the connector with the large pin (pin 1 - ground)
on the SBE 63 sensor mount. Remove any trapped air by
burping
or
gently squeezing the connector near the top and moving your fingers
toward the end cap.
OR
MCBH Connector
– Install the cable connector, aligning the pins.
C.
Place the locking sleeve over the connector. Tighten the locking
sleeve finger tight only.
Do not overtighten the locking sleeve and
do not use a wrench or pliers.
4.
Connect the I/O cable to the 4-pin RS-232 auxiliary sensor connector on
the CTD, using the same technique as in Step 3 (see the CTD manual for
the end cap detail).
5.
Verify that the hardware and external fittings are secure.
The SBE 63 is ready to go into the water.
Integration with MicroCAT CTD
(SBE 37-SMP-ODO, 37-IMP-ODO, 37-SIP-ODO)
The SBE 63 is factory-integrated with the ODO MicroCAT.
You can communicate with the SBE 63
through
the MicroCAT with
Send63=command
;
command
can be any command recognized by the
SBE 63. Verify that the SBE 63 is set up as follows by sending
Send63=GetSD
(see
Command Descriptions
above for command details):
•
SetBaud=2400
(factory set; cannot be changed by command through the
MicroCAT)
•
SetEcho=1
•
SetAvg=1 to 16; recommended value is 2
•
SetFormat=1
•
SetAutoRun=0
For use with the SBE 37-IMP-ODO, 37-SMP-ODO with RS-485 interface,
and 37-SIP-ODO with RS-485 interface, verify that <SerPause> in the
GetSD
or
GetHD
response is 1 (factory set). For the other MicroCATs, <SerPause>
can be set to 0 or 1.
CAUTION:
Do not use WD-40
or other petroleum-
based lubricants, as they will damage
the connector.
I/O cable
Locking sleeve
Note:
Precede
Send63=
with
#ii
(ii=
MicroCAT ID) for the 37-IMP-ODO
or the RS-485 versions of the
37-SMP-ODO and 37-SIP-ODO
(for example,
#01Send63=GetSD
).