Manual revision 005
Section 4: Deploying and Operating SBE 63
SBE 63
19
Example 2: Autonomous Sampling Setup for SetAutoRun=1
for use
with power supply and logger.
Set up to start sampling when power is applied (
SetAutoRun=1
).
Set up to echo commands (
SetEcho=1
), take samples at 10-sec intervals
(
SetInterval=10
), average 2 measurements per output sample
(
SetAvg=2
), output converted oxygen and temperature data
(
SetFormat=0
), and set the delay before sampling to the minimum value
(
SetBootDelay=
). Verify setup with status (
GetHD
) command. Remove
power.
(Apply power, then select Connect in Seaterm232’s Communications
menu to connect and wake up.)
S>
SETAUTORUN=1
S>SETECHO=1
S>
SETINTERVAL=10
S>
SETAVG=2
S>
SETFORMAT=0
S>
SETBOOTDELAY=1
S>
GETHD
(to verify setup)
(Remove power from SBE 63.)
When ready to begin sampling, connect SBE 63 to power supply and data
logger.
(Apply power – SBE 63 automatically begins sampling at 10-sec
intervals.)
When ready to stop sampling:
(remove power.)