Service manual - X10
19
Troubleshooting
6.1.
General diagnostics and faultfinding
Problem
Cause
Solution
Wheelchair does not switch
on
Bus cables not connected
Check bus cable
connections of the joystick
module to the power
module (can also be
through the actuator
module!)
Pins in connectors are
loose or damaged
Check connectors for loose
or damaged pins. If
damaged replace cable.
Bus cable defect
Check all bus cables for
damage. Bypass each
bus cable with new one
to check functionality.
Fuse defect / Thermal trip
Check fuses and replace if
needed.
Battery connections are
loose
Check battery connections
and restore if needed.
Battery voltage too low
Measure battery voltage
and charge batteries.
Battery defect
Measure battery voltage.
Voltage below 10 V can
indicate that one battery is
defect. Replace batteries.
Remote control or button
defect
Check by replacing with
new one.
Power Module defect
Check by replacing with
new one.
Wheelchair continuously
drives slower than normal
Speed limit due to
micro switch signal
from seating function.
Check if all seating options
are in neutral position.
Measure resistance from
the speed limit cable in
different seating positions.
Bad cable connections or a
broken resistor influences
the drive inhibit signal.
Incorrect programming
Check for correct program.
Re-program if needed.
Содержание X10
Страница 1: ...Service Manual X10 V1 0 from March 1st 2017 ...
Страница 38: ...38 Service manual X10 Cable and module schemes 7 4 R Net controls ...
Страница 39: ...Service manual X10 39 Cable and module schemes ...
Страница 40: ...40 Service manual X10 Cable and module schemes ...
Страница 41: ...Service manual X10 41 Cable and module schemes 8 Version history Version Release date Changes ...
Страница 42: ...Appendices 42 Service manual X10 9 Appendices ...
Страница 43: ...Parts X10 V1 0 from 01 03 2017 ...
Страница 45: ...3 ...
Страница 46: ...4 ...
Страница 48: ...TABLE OF CONTENTS X10 EN NP 6 ...
Страница 79: ......
Страница 80: ......