CHAPTER 4: OPERATION
087-0038 REV L
SCOTT SAFETY
9
Cal Due – Indicates when next calibration is due in days.
Warning:
Operating a device that has exceeded its calibration date can cause false
readings of detected gases. Do not trust invalid readings obtained while device is out
of calibration. Operating a device that is out of calibration may not accurately
indicate environmental conditions which can lead to death or injury.
Warning:
If CAL EXPIRED is displayed upon power on start up sequence indicating
the device does not have valid current calibration, press the Right Operation Button to
override and immediately calibrate the device. A device that has not been properly
calibrated can lead to death or injury.
Warning:
Never operate the device in gas monitoring mode with the calibration
adapter attached. This can cause false readings of detected gases. Failure to do so
can lead to death or injury.
After this sequence is complete, the device resumes typical monitoring.
Содержание Protege
Страница 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 2: ......
Страница 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Страница 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Страница 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Страница 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Страница 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Страница 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Страница 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Страница 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Страница 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Страница 103: ......
Страница 104: ...Monroe Corporate Center P O Box 569 Monroe NC 28111 Telephone 800 247 7257 FAX 704 291 8330 Web www scottsafety com ...