PAGE 27
(5) Secure and install the new membrane set
on the probe properly.
After instaIIing the new membrane set with
electrolyte fil led, it is normal to found some
small bubbles inside of the membrane.
(6)
Proceed “calibration” as above section after
replace the membrane set.
NOTE: Under normal use, the extra
membranes set enclosed in package will be enough to cover the replacement
need for entire DO probe life time.
The pH glass bulb should always be moistened by using the probe cover with storage
buffer. Always rinse the pH electrode with pure water or distilled water before use.
Never touch or rub the glass bulb for lasting pH electrode life.
Please soak the conductivity probe in distilled water for 30 min before each usage to
avoid the inert effect.
Only use soft brush or clean water to clean probe surface if dirt is attached on surface.
If the electrode surface is contaminated, place the probe in a diluted detergent for
about 15 minutes, then rinse it with distilled water. After clearing, please keep probe
dry for storage.
PH probe maintenance
COND probe maintenance