PAGE 26
4. After scrap off the white residues, use clean water to wash away the remaining
impurities.
HOW TO REPLACE MEMBRANE SET
a) Please replace membrane set when any of following situation occurs:
b) The probe has been used for >1 year
c) Seeing error code E2/E3 but the probe central working area is still clean.
d) Electrolyte is less than 1/2 fulI
e) Membrane is broken or wrinkled
f) Measuring speed is much slower than before.
The procedures are:
(1) Prepare a new membrane set.
NOTE: Due to the structure of membrane is delicate, once the membrane set has
been removed from the probe, it is “not suggested” to re-install ed the same
membrane set repeatedly as it might
influence the reading.
(2) Rotate to unscrew the probe protective
cover and remove the membrane set from
probe.
(3) Follow the steps in above section to
“clean the electrode”.
(4) Slowly pour the electrolyte solution to the
new membrane set till full.