P60 is used to review and customize the suitable setting for D.O. Measuring.
For example, if the measured target is sea water or in plateau, it is required to setup
correct salinity value and altitude.
1. Enter to setup mode and press
△
or
▽
key in the setup mode to enter COE P60
screen (Fig 10). Press ENTER key to display P61 and review previous DO
calibration temperature (Fig 11). This value helps to judge root cause while
seeing error code E21 on LCD. E21 means the measured temp. & calibration
temperature has 10 degree C difference.
2.
Then press “ENTER” into P62 ppt, the value will be flashed on the screen (Fig
12). You can manually enter the salinity compensation. (Use salinity value
measured by the conductivity probe). Press
△
or
▽
adjust the salinity value. The
value can be adjusted from 0.0 to 42.00ppt.
3. Press Enter to confirm. then enter P63 M to program the height from sea level
(Fig 13). The value will be flashed on the screen. You can manually press
△
or
▽
key to adjust from 0-3500M (100M per step), press ENTER to confirm and enter
COE P64 FAS screen.
4. See page 8 for the definition of FAS and SLOW. You can press
△
or
▽
key to
adjust, press ENTER to confirm.
P60 DO Review and Set parameters adjustable