PAGE 18
P50 is used to customize the factor to convert conductivity into TDS value.
1. Enter to setup mode and press
△
or
▽
key in the setup mode to check td F P50
display.
2. Press ENTER key to enter and press
△
or
▽
to adjust. The default is 0.50 and
adjustable range is 0.30 -1.00. Press ENTER key again to save and return to
P50 screen.
This function allows you to select the unit of the temperature:
1. Press
△
or
▽
key in the setup mode to enter the temperature unit P70 screen.
Press the ENTER key to select temperature unit setting.
“unt” will display on the
left side of the screen (Fig 14)
2. At P70 screen, press ENTER key to enter P71. The default "C" will be flashed on
the main screen. (Fig 15). If the unit you need is degree
o
C, press ENTER key to
confirm.
3. If the unit you desired is degree
o
F, press
△
or
▽
key to change the C to F and
then press ENTER key to confirm. The meter will return to P70 screen (Fig 14)
P50 Cond. TDS factor (Adjustable)
P70 Temperature unit adjustable
Содержание SP-850048
Страница 1: ...PAGE 1 DO Multi Parameter Meter SP 850048 Instruction Manual ...
Страница 15: ...PAGE 15 ...