3.3.7 OUTP RESU
Touch the corresponding zone. The following items will be displayed in the soft key zone.
LEVEL
It is used for setting the OUTP RESU as LEVEL.
PULSE
It is used for setting the OUTP RESU as PULSE.
Touch the corresponding key in soft key zone to select the relevant function.
3.3.8 PULSE WID
This function is used to set the pulse width of pulse signal for Handler PICKOUT (
this
item can be changed when the OUTP RESU is set as PULSE
). The setting range is
1-25ms.
Operating Procedures for setting the PULSE WID:
Procedure-1:
1. Touch this key to select this option and then the colour of this option becomes the
colour of the cursor.
2. Touch this key (cursor) again to enter into the number soft keyboard, input the
required pulse width and then press the ENTER key to confirm and return to
<PICKOUT SETUP> page.
Procedure-2:
Touch the average zone needs to be modified on the screen and the following items will
be displayed in the soft key zone.
↑↑
(
(
(
(
+ +
)
)
)
)
Touching this key increases the pulse width (set value) rapidly by 5.
↑
(
(
(
(
+
)
)
)
)
Touching this key increases the pulse width (set value) by 1.
↓
(
(
(
(
-
)
)
)
)
Touching this key decreases the pulse width (set value) by 1.
↓↓
(
(
(
(
- -
)
)
)
)
Touching this key decreases the pulse width (set value) rapidly by 5.
3.4 Start Test
3.4.1 Test Method
Please connect the DUT (especially the DUT with polarity, such as the capacitor)
according to the method as shown in figure 3-4 before testing.