SME1202A
5)
Performance Index
5.1 Measurement Function
5.1.1 Measurement Parameters and Notations
R
:
Resistance
I: Leakage current
5.1.2 Range
Resistance Mode: AUTO, MANU (HOLD, UP, DOWN)
5.1.3 Trigger
Manual, external, bus
Manual : Press the “TRIGGER” button on the panel, the instrument will test a DUT once
and display the result. This mode keeps in waiting mode when it is not used.
External: Test a DUT once and display the result when the instrument receives an external
“start up” signal from the footswitch of HANDLER interface on the rear panel.
BUS : The measurement of the instrument will be triggered through the communication
interface.
5.1.4 Mode of Test Terminal
2-terminal measurement mode
HV (-): DCV output terminal
INPUT: Sampling input terminal
5.1.5 Measurement Speed (when MEAS DISP & BIN DISP is set as OFF)
Speed
Parameter
FAST
SLOW
R
Meas/sec
≤
30ms
Meas/sec
≤
60ms
I
Meas/sec
≤
30ms
Meas/sec
≤
60ms
5.1.6 Display Digits
Item
Resistance
4-digit significant figures
Leakage
current
Range 1
:
100nA
Range 2
:
10nA
Range 3
:
1nA
Range 4
:
100pA
Range 5
:
10pA
Range 6
:
1pA
Time
0.1s