2 Commissioning
Page 9
STAT
MATIC
smart 1,2,3
EN
2
Commissioning
Environmental conditions
There are a number of factors that can influence the performance of your STAT
MATIC
smart system. Please
consider these factors carefully and choose a suitable location for installing the device.
•
Temperature and air humidity
Do not set up your STAT
MATIC
smart system at a location where it is exposed to direct sunlight or a heat source
(for example a heater). The recommended operating temperature lies between 5 and 35°C with air humidity
between 25 and 85%.
•
Environment
The STAT
MATIC
smart should be operated in a clean, dust-free and well-ventilated environment.
•
Work surface
The STAT
MATIC
smart should be installed on a flat, level and water-resistant as well as oil-resistant surface. Do
not install or operate your unit on a sloped surface.
• Electromagnetic environment
The STAT
MATIC
smart has been tested for electromagnetic emissions and meets the decisive standards. While
the product itself emits no radiation, it can be influenced by other radiating devices. We therefore recommend that
you keep the device away from potential sources of interference.
Electrical Requirements
DANGER! Mortal danger with improper handling of power sources! Danger of electrical shock
•
For connection to the mains, use a properly grounded and secure socket providing the stated voltage (see type
plate on the back of the STAT
MATIC
smart).
•
Avoid using multiple sockets. When using power bars with surge protection, always connect only a single device.
•
The mains input connector on the back of the device is to be used as a separator from the power supply.
Media connections
Plug the mains plug into the socket.
Attach the compressed air hose to the coupling and tighten.
The set pressure should be between
4.5 bar
and 4.3 bar!
In the event of a deviating display, pull out the grip for regulating pressure and turn it to the right or the left until
the manometer shows 4.5 bar.
Push the grip back into the pressure regulator and fixate it.
In order to ensure a safe and trouble-free operation, only use compressed air in accordance with the information
in the "Technical Data" section.