3 Operation
Page 15
STAT
MATIC
smart 1,2,3
EN
"Special instrument" program
Special instruments can be maintained and cleaned using this program (such as prophylaxis). The times require for
maintenance and cleaning can be individually adjusted by a service technician.
Turn on the mains switch.
LED
(Stand-by) lights up green. "Cl Maintenance" will appear on the display
Open the front door.
Place the instruments on the adapters provided for this purpose (instruments must snap into place). A frame will
appear on the display around the station number per occupied slot.
Close the front cover.
Press the button 3x. "Special instrument" will appear on the display.
Push the
(Program Start) button on the control membrane keyboard.
Program runs automatically. During the "Cleaning" program, a clock will count backwards until 00:00 on the display
All attached instruments are cleaned and maintained for inside in turn. After a successful process, the program slots
will be displayed inversely. The display shows: "Special instrument successful".
Open the front door.
Remove regular and contra-angle handpieces by pressing the locking button.
Pull off turbines downwards (press support lever).
The frame around the slot number will disappear from the display.
•
Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using
disinfectants listed by the VAH (such as SciCan OPT
IM
BLUE).
•
Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's
instructions on reprocessing (such as disinfection/sterilization in the STAT
IM
).
•
Always perform a function test before using the instruments on patients. Some liquid may come out.