5 Troubleshooting
Page 21
STAT
MATIC
smart 1,2,3
EN
5
Troubleshooting
Error
Display
Acoustic signal
Cause
Correction
LEDs don't light up
No display
–
No power supply
Activate mains switch
Plug in the power supply
plug
Defective fuse
Replace the fuse
(see Chapter 6
"Troubleshooting")
Compressed air
ATTENTION
Compressed air error
Fault LED (L2) flashes
Interval
– Beep
0.5 seconds on,
8 seconds off;
10 times
Compressed air hose
not connected
Connect compressed air
hose
Compressed air hose
pinched
Remove flow disruption
Mains supply defective Create supply
Pressure reducer
defective
Please consult the SciCan
customer service.
Pressure switch
defective
Please consult the SciCan
customer service.
If available; filter in
supply line full
Replace the filter
Top cover open
WARNING
Close the top
cover
Fault LED (L2) flashes
–
Top cover not closed
correctly
Close top cover tightly
Magnet on top cover is
missing
Please consult the SciCan
customer service.
Program doesn't start
ATTENTION
Close the door
–
Front doors are not
correctly closed
Close front doors tightly
ATTENTION
No slots
occupied
–
No instruments in the
device
Attach the instruments to
the couplings
Switching flap does
not recognize
instrument
Turn the instrument until a
frame is displayed around
the station number
Process interrupted
ATTENTION
Process interrupted
–
Front doors are not
correctly closed
Close front doors tightly.
Instrument is not
attached to the
coupling correctly
Push the instrument firmly
onto the coupling
Undefined fault
ATTENTION
Please consult the
customer service.
Interval
– Beep
2 seconds on,
4 seconds off
Internal device fault
Please consult the SciCan
customer service.
Memory full
ATTENTION
Not enough memory
available
–
Internal memory full
Export log data to USB stick