6. Troubleshooting Cont’d.
Excessive steam issuing from the
front of the machine.
Message REFILL RESERVOIR,
Machine will not start.
The level of the water in the reservoir is low. Refill the
reservoir. Refer to the steps described in Section 3.4
Filling the Reservoir.
Message PRINTER FAULT displayed on
LCD Printer is not printing.
Check for a paper jam. If the paper is jammed, follow
the paper removal procedures outlined in the following
section. Power unit
OFF
for 10 seconds and then
power
ON
again. If the paper is still jammed, follow
the disassembly procedures outlined in Section 7.5
Removing Paper Jams.
The printer does not work.
Make sure that the printer cable is connected securely
with the connector on the back of the Statim 2000 and
the Statprinter.
Make sure that the printer is powered
ON
. Power unit
OFF
for 10 seconds and then power
ON
again.
The printer appears to work, but nothing is
printed on the paper.
Ensure that the paper is loaded properly (refer to
Section 7.4 Installing paper into the Statprinter).
Check to see that the paper leaves the paper roll from
the top of the roll. This means that the treated surface
of the coated thermal paper will be in contact with the
thermal print head.
Time and date are incorrect.
The time and date have not been set. See Section 7.3
Setting the Time and Date.
Message WATER QUALITY IS NOT
ACCEPTABLE. Machine will not start.
You have used water which is not steam-process
distilled or is improperly distilled.
Empty the reservoir and refill with steam-process
distilled water containing less than 5ppm total disolved
solids (having conductivity of less than 10 µS/cm).
If you have the water conductivity meter, check the
quality of the water before refilling the reservoir. Refer
to the steps described in Section 3.8 Shipping the Unit
to empty the reservoir.
Remove and reinsert the cassette. Attempt another
cycle. If the problem persists remove and check the
cassette seal for misalignment or damage. Replace
the seal if required. Be careful as the metal parts will
be hot and the cassette will contain hot steam.
If the leak persists, turn the unit
OFF
, remove and
unload the cassette and contact your SciCan dealer.