4. Instructions For Use Cont’d.
For detailed instructions on how to handle, use and dispose of both the biological and
chemical indicators, please consult the product literature accompanying the 3M Attest
TM
biological indicators or contact the manufacturer directly.
To use the indicators with the Statim, follow these steps:
1. Place the appropriate biological indicator in the Statim chamber.
2. Process the load in the sterilizer according to your usual practice.
3. Ensure that the message “Cycle Complete” is displayed on the LCD after the cycle
is finished.
4. Recover the biological and / or chemical indicator and process further according
to the literature that accompanied the indicator.
At the first indication of a potential sterilization failure:
1. Do not process any more instruments until favourable test results have
been returned.
2. Ensure the correct indicator type was chosen.
3. Ensure the cassette was not overloaded. Consult the earlier portion of this section
for proper loading instructions.
4. If the results do not change, do not process any more instruments within the
Statim and contact your SciCan dealer for further assistance.
Because the turnaround time for the 3M Attest
TM
is up to 48 hours, it is recommended
that the tests be conducted so that the incubation period occurs during a period of
planned downtime such a the last cycle before a weekend.
4.5 Instrument Mass Guide
Instrument
Scissors
Dental scalers
Forceps
Dental handpiece
Wrapped Instrument rack
Unwrapped instrument rack
Suction cannula
Plastic mouth mirror
Impression tray
Plastic x-ray positioning ring
Typical Instrument Mass
30 g (0.96 ozt)
20 g (0.64 ozt)
15 g (0.48 ozt)
40 to 60 g (1.3 ozt to 1.9 ozt)
260 g (8.36 ozt)
225 g (7.23 ozt)
10 g (0.32 ozt)
8 g (0.3 ozt)
15 to 45 g (0.48 ozt to 1.5 ozt)
20 g (0.64 ozt)
NOTE
: The above weights are to be used as reference only. For exact weights of your
instruments, consult the manufacturer’s specifications.