3
Operation
3.3
ECG Settings
A"#$0$
Page 18
A
rt
.-
no
.:
2
.5
4
0
0
37
R
ev
.:
a
3.3.1
Automatic Format 1 and 2 Internal Printer
Press the SELECT softkey to choose from the following options:
Two separate Auto formats can be defined for the internal printer.
General
Lead
Interpretation
Filter
Auto. Format 2 external
ECG Settings
Auto Format 2 internal
external
ECG Settings
Automatic Format 1
internal
Auto. Format 1 external
NEXT
Parameter
Options
Description
ECG Printout
No Printout
No printout of the ECG given at the end of an auto mode recording (the
recording can be stored in the memory and printed at a later time if re-
quired).
Short at 25 mm/s
Leads are printed in short form (1 sheet/lead group) at 25 mm/s.
Short at 50 mm/s
Leads are printed in short form (1 sheet/lead group) at 50 mm/s.
Long at 25 mm/s
Leads are printed in long form (2 sheets/lead group) at 25 mm/s.
Long at 50 mm/s
Leads are printed in long form (2 sheets/lead group) at 50 mm/s.
Average Cycles
No Printout
No printout of the ECG given at the end of an auto mode recording (the
recording can be stored in the memory and printed at a later time if re-
quired).
4x3, 25 mm/s
Leads are printed in a 4 x 3 format at 25mm/s
4x3, 50 mm/s
Leads are printed in a 4 x 3 format at 50mm/s
2x6, 50 mm/s + 1 Rhythm
Leads are averaged over the entire 10 second recording and printed in
2 groups of 6 at 50mm/s, with the one rhythm lead at the bottom of the
page.
1x12, 50mm/s + 2 Rhythm
Leads are printed in a 1 x 12 format at 25mm/s - with 2 rhythm leads
printed.
Rhythm Lead 1
I, II, III, aVR, aVl, aVF, V1 to V6 Select any lead.
Rhythm Lead 2
I, II, III, aVR, aVl, aVF, V1 to V6 Select any lead.
Measurements
On/Off
Select on or off to print a detailed table of measurement results.
Markings
On/Off
Select on or off to print reference markings on the ECG average cycle
print. A vertical marker shows the beginning and end of P wave and QRS,
and the end of the T wave
Interpretation
On /Off
Select on or off to print interpretation statement
The sensitivity of the interpretation and the printing/not printing of abnormality state-
ments can additionally be set for the interpretation. These are described in paragraph
Содержание AT-101
Страница 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Страница 6: ...Page 4 Art no 2 540037 Rev a ...
Страница 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Страница 58: ...7 Drawing and spare parts 7 1 Drawings Printer A 0 Page 56 Art no 2 540037 Rev a Fig 7 4 Detail cabling ...
Страница 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Страница 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...
Страница 66: ...9 Index A 0 Page 64 Art no 2 540037 Rev a ...