8
Appendix
8.1
Checklist
A"#$0$
Page 60
A
rt
.-
no
.:
2
.5
4
0
0
37
R
ev
.:
a
3. The mains indicator lamp (6) switches off and the symbol battery is dis-
1. The paper has to stop exactly at the perforation. If this is not the case, con-
trol the paper mark comparison.
5.2.8 Printer quality test page 36
1. Check the blackness for regularity and good readability on the complete
2. Check on the 25 mm/s printout if the space between 2 R peaks according
to the paper grids. The space should be 25 mm and may not show a dif-
ference of more than ± 0.5 mm.
1. Check on the 25 mm/s printout if the space between 2 R peaks according
to the paper grids. The space should be 25 mm and may not show a dif-
ference of more than ± 0.5 mm.
HR is set to 60
5.2.10 Parallelism test page 37
1. All calibration impulses should be lined up vertically and exactly below
2. The maximum deviation may not be more than ± 0.5 square. If the values
are outside this tolerance, the mechanical adjustment of the print head
has to be corrected.
1. Check the amplitude height of the calibration impulse according to the pa-
per grid. It should be 20 mm ±0.5 square.
5.2.12 ECG lead and patient cable test page 38
1. all leads (1) are blinking
2. Uel for all leads is -350 to -550 mV
3. all leads stops blinking
4. Uel for all leads is -15 ...+15 mV
5. Udif = 3985 ...4015 mV
6. TPH = 18 ...30 °C
7. Check amplitude and polarity according printouts
5.2.13 External printer test page 41
1. all leads (1) are blinking
2. Uel for all leads is -350 to -550 mV
3. all leads stops blinking
4. Uel for all leads is -15 ...+15 mV
5. Udif = 3985 ...4015 mV
6. TPH = 18 ...30 °C
7. Check amplitude and polarity according printouts
5.2.14 Communication (RS-232) Test page 41
1. The screen shows hex characters when the serial port is ok.
1. The safety test is carried out in accordance with the EN 60601-1, Clause
Add protocol to the checklist
Reference/Testpoints
OK
False Remark
Содержание AT-101
Страница 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Страница 6: ...Page 4 Art no 2 540037 Rev a ...
Страница 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Страница 58: ...7 Drawing and spare parts 7 1 Drawings Printer A 0 Page 56 Art no 2 540037 Rev a Fig 7 4 Detail cabling ...
Страница 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Страница 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...