5
Care & Maintenance
5.3
Safety Tests
A"#$0$
Page 42
A
rt
.-
no
.:
2
.5
4
0
0
37
R
ev
.:
a
5.3 Safety Tests
The safety test is carried out in accordance with the EN 60601-1, Clause 18 and 19.
This test may only be carried out with a tester that fulfils the above mentioned norms
and has been calibrated in accordance with ISO norms.
To carry out the test, follow the instructions of the manufacturers.
Documentation
Note the results or have them printed by the tester. Always include one copy of the
results with the repair report. The original remains with the device and is given to the
customer for his files.
5.3.1
Maximum values safety test
Ground Resistance:
!
0.2
"
Voltage
Type BF
Type CF
normal condition
first error
normal condition
first error
Earth current general [mA]
0.5
1.0
0.5
0.5
Shell current [mA]
0.1
0.5
0.1
0.5
Patient current [mA]
0.1
0.5
0.01
0.05
Patient current [mA]
(Mains voltage at signal entrance and exit)
--
--
--
--
Patient current [mA] (mains voltage at used part)
--
5.0
--
0.05
Patient independent current [mA]
Direct Alternating Current [mA]
0.01
0.1
0.05
0.5
0.01
0.01
0.05
0.05
Содержание AT-101
Страница 1: ...A 0 ervi e andboo3 Art no 7 5400 Rev a 400 A 0 7 annel E G Devi e ...
Страница 6: ...Page 4 Art no 2 540037 Rev a ...
Страница 42: ...5 Care Maintenance 5 2 Functional test A 0 Page 40 Art no 2 540037 Rev a ...
Страница 58: ...7 Drawing and spare parts 7 1 Drawings Printer A 0 Page 56 Art no 2 540037 Rev a Fig 7 4 Detail cabling ...
Страница 63: ...Page 61 Appendix 8 A 0 Service Handbook Checklist 8 1 Art no 2 540037 Rev a Other remarks ...
Страница 64: ...8 Appendix 8 1 Checklist A 0 Page 62 Art no 2 540037 Rev a ...
Страница 66: ...9 Index A 0 Page 64 Art no 2 540037 Rev a ...