Further hygiene measures
Abb. 4
Abb. 5
Cover the central section and
turn the vaporizer on. The
water will have evaporated
after approximately 20 minutes
and the disinfection process
will be concluded.
In the home:
After sealing the socket with
the stopper the hand unit and
individual parts can be cleaned
in a dishwasher (open chamber
side facing downwards).
In clinics:
After sealing the socket with
the stopper the hand unit can
be disinfected / sterilized
through immersion disinfec-
tion, placing it in a chemical-
thermal disinfector (60°C for
10 minutes) or in a gas steri-
lizer.
The nebuliser head, valves,
mouthpiece, baffle plate, sea-
ling ring and filter holder can
be disinfected, gas-sterilized
or autoclaved at up to 134°C
as described above.
After cleaning and disinfection of the unit in a vaporizer / dishwasher or after chemical-thermal disinfec-
tion the unit must be allowed to cool for approximately 30 minutes before removing the parts and start-
ing the next inhalation procedure.