• Never use a new battery pack
until it has been fully charged.
A new battery pack must be
charged for a minimum of 4 –
5 hours.
• The hand unit cannot be oper-
ated by the battery pack
whilst it is being recharged.
The battery pack has been spe-
cially developed for use with
the multisonic inhalation unit
and is designed for an inhala-
tion period of at least 60 min-
utes.
The initial charge is very impor-
tant for the performance of the
battery pack.
The optimal operating time is
reached after 2 – 3 complete
charges. If used and charged
correctly, the battery pack can
be charged at least 500 times.
As with all Ni-MH batteries, the
battery pack will only deliver
its full power if it is fully
charged.
If the battery pack is mistreat-
ed then its capacity will be re-
duced. It should not be stored
at a temperature > 40 ºC.
When not in use the battery
pack should be stored in a cool
dry place.
21
The battery pack must then be
recharged up to a capacity of
> 20% (LED 3 lit up) before it
can be connected to the nebu-
liser unit.
The capacity display (LED 3 to
LED 7) will always be active
when the supply voltage is
connected to the socket X1, or
if the nebuliser unit is connec-
ted to it.
The display may be activated
by pressing the button T1.