Digital CCD Line Scan Camera SK 10680 DJR ( Rev.1.4. / 20.09.2005 ) - Manual
Page
7
9. Gain / Offset - Settings
P1
P2
Camera backside
To adjust the gain / offset settings the camera does
not need to be opened. The trimmers for balancing
P1 and P2 are accessible from the outside.
The camera is shiped with optimum Gain / Offset
calibration. If a recalibration becomes necessary
please follow the instructions given below:
1. Adjust the individual offset of the line sensor video
signal to zero voltage (‘00’ digital) using P1.
2. Use P2 to adjust the maximum output voltage.
Illuminate the camera sufficiently.
The maximum output voltage is set to ca. 2.5 volts
standard (‘FF’ digital).
Camera board,
sensor side
The spectral range of the camera can be select by
soldering a bridge over the appropriate pads.
B
R
G
Although this manual has been reviewed carefully for
technical accuracy, errors are possible. The reader is
kindly asked to contact us, if errors are suspected.
The indicated circuits, descriptions and tables are not
warranted to be free from rights of third parties.
With the statements in the technical descriptions only
assembly groups are specified. Characteristics as well
as the suitability for a particular purpose is not guaran-
teed.
The warranty period for the CCD line scan camera is 24
months. The warranty ends with inappropriate actions.
10. References and Warranty
EC-Declaration of Conformity
This product meets the requirement
of the EC directive 89/336/E.E.G.
The requirements of DIN EN 61326
are fullfilled.