                 TC-BA-e-1712 

Resolution: 0.1 µm (0 to 99.9 µm) 
                   1 µm (over 100 µm) 
 
Accuracy: 
- Standard:   3% of the measured value or ± 2.5 µm or 0.1mil 
This is valid within a tolerance range of ± 100 

μm around the individually measured 

range, if a two-point calibration was performed within this tolerance range. 
-- 

Off-Set Accur: 

1 % of the measured value or Min ± 1.0 µm 

                This is valid within ± 50 

μm around the 

Off-Set Accur

 point. 

 
PC- interface: with RS-232C interface 
 
Power supply:   4x 1.5V AAA (UM-4) battery 
 
Operating conditions:   Temperature: 0 to 50°C 
                                     Humidity:       less than 80%                                          
                                                                          
Dimensions:   127 x 67 x 28mm (5.0 x 2.6 x 1.1Inch) 
 
Weight:   about 81g (batteries not included) 
 
Scope of delivery:  - Carrying case 
                              - Instruction manual 
 
                            According to types: 1. Incorporated sensor at F model: 
                                                               TC 1250-0.1 F, with button F/N 

                                                           2. Incorporated sensor at N model: 

                                                               TC 1250-0.1 N, with button F/N 
                                                            3. Incorporated sensor at FN model: 
                                                               TC 1250-0.1 FN, without button F/N u. S/C  
 
                            - Adjustment foils, at every model  
                            - Zeroing plate (Aluminum) at N model             
                            - Zeroing plate (iron) at F model 
                            - 2 Zeroing plates (Aluminum and iron) at FN combination model 
                                                     
Optional accessory: Cable & software for RS-232C:  ATC-01 

 
 
 
 

Содержание TC 1250-0.1 F

Страница 1: ...n E Mail info sauter eu Tel 49 0 7433 9933 199 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version 1 2 08 2017 GB Illustration TC 1250 0 1FN TC BA e 1712 PROFESSIONAL MEASURING ...

Страница 2: ...ge If you have any queries wishes or helpful suggestions do not hesitate to call our service number Models available TC 1250 0 1 F TC 1250 0 1 N TC 1250 0 1 FN TC_car 1250 0 1FN separate instruction manual available Summarize 1 Features 3 2 Specifications 3 3 Front Panel description 5 4 Measuring procedure 5 5 Zero Adjustment Calibration 6 6 Battery replacement 7 7 Adjustment foils 7 8 General adv...

Страница 3: ...d on magnetic metals e g steel iron nickel etc The N sensor measures the thickness of non magnetic coatings on non magnetic metals It is used on anodizing varnish paint enamel plastic coatings powder etc It can be applied on non magnetic materials like aluminium brass stainless steel etc The N sensor measures the thickness of non magnetic coatings on non magnetic metals It is used on anodization v...

Страница 4: ...V AAA UM 4 battery Operating conditions Temperature 0 to 50 C Humidity less than 80 Dimensions 127 x 67 x 28mm 5 0 x 2 6 x 1 1Inch Weight about 81g batteries not included Scope of delivery Carrying case Instruction manual According to types 1 Incorporated sensor at F model TC 1250 0 1 F with button F N 2 Incorporated sensor at N model TC 1250 0 1 N with button F N 3 Incorporated sensor at FN model...

Страница 5: ...ymbol Fe F for magnetic metals or NFe N for non magnetic metals which is indicated on the display It switches to the automatic measurement mode which also correctly identifies the zeroing plate or any other base material b The sensor 3 1 has to be placed onto a coating layer to be measured The reading on the display is the thickness of the coating layer This can be corrected by pressing the Plus k...

Страница 6: ...s the Power key 3 6 5 Zero Adjustment Calibration a Zero adjustment for Fe F and NFe should be carried out separately The Zero plate of iron has to be chosen if Fe is displayed and zero plate of alumi num has to be chosen if NFe is displayed Sensor 3 1 has to be placed carefully onto the appropriate base plate or another un coated base material The Zero key 3 3 has to be pressed without lifting th...

Страница 7: ...ise a The instrument should always be calibrated on the uncoated base material to be measured instead of the base plate included in the delivery Then accuracy is more precise right from the start b Sensors will eventually wear off Life of the sensor will depend on the number of measurements taken and how abrasive the coating is The change of a sensor should only be performed by qualified persons 9...

Страница 8: ...y be adjusted by professional persons Generally said The bigger the value of Ln the smaller the reading on the display for the same coat ing thickness Only a small change of the value of LN causes a big change in the reading of the upper measurement range at 500µm 20 mil The value of LN has to be adjusted as follows By pressing the Power key It lasts about 11 seconds from starting to press this ke...