6 Battery replacement
a) If the battery symbol ``+/-`` appears on the display, the batteries should be re-
placed.
b) Battery cover 3-8 has to be removed and batteries have to be taken out.
c) Batteries (4x1.5V AAA/UM-4) have to be installed correctly into the case. Please
pay attention to polarity!
d) If the instrument is not going to be used for an extended period, batteries have to
be removed.
7 Adjustment foils
This instrument includes in its scope of delivery a kit of adjustment foils with several
foils of different measurement ranges. But there is always covered a measurement
range between 20 and 2000µm. They are also optional available as a spare part, ar-
ticle number ATB-US07.
8 General advise
a) The instrument should always be calibrated on the uncoated base material to be
measured instead of the base plate included in the delivery. Then accuracy is more
precise right from the start.
b) Sensors will eventually wear off. Life of the sensor will depend on the number of
measurements taken and how abrasive the coating is. The change of a sensor
should only be performed by qualified persons.
9 How to restore Factory Settings
a) In the following cases it is recommended to restore factory settings:
- The instrument does not measure any more.
- Measurement accuracy is degraded caused by the abraded sensor or affected by
environmental conditions.
- After the replacement of a new sensor.
b) Procedure:
Factory setting is including both, the settings for “Fe” (F) and for “NFe” (N).
Please observe which symbol is shown on the display.
It can be only performed one setting after another. The procedure is as follows: