TC-BA-e-1712 

 

 

6  Battery replacement 

 

a)  If the battery symbol ``+/-`` appears on the display, the  batteries should be re-
placed. 
 
b) Battery cover 3-8 has to be removed and batteries have to be taken out.       
 
c)  Batteries (4x1.5V AAA/UM-4)  have  to be installed  correctly into the case. Please 
pay attention to polarity! 
 
d) If the instrument is not going to be used for an extended period, batteries have to 
be removed.       

7  Adjustment foils 

 
This instrument includes in its scope of delivery a kit of adjustment foils with several 
foils of different measurement ranges. But there is always covered a measurement 
range between 20 and 2000µm. They are also optional available as a spare part, ar-
ticle number ATB-US07. 

8  General advise 

 
a) The instrument should always be calibrated on the uncoated base material to be 
measured instead of the base plate included in the delivery. Then accuracy is more 
precise right from the start.        
   
b) Sensors will eventually wear off. Life of the sensor will depend on the number of 
measurements taken and  how abrasive the coating is.  The change of a sensor 
should only be performed by qualified persons.      

9  How to restore Factory Settings 

 
a) In the following cases it is recommended to restore factory settings: 
- The instrument does not measure any more. 
-  Measurement accuracy is degraded  caused by the abraded sensor or affected by 
environmental conditions. 
- After the replacement of a new sensor. 
 
b) Procedure: 
Factory setting is including both, the settings for “Fe” (F) and for “NFe” (N). 
Please observe which symbol is shown on the display. 
It can be only performed one setting after another. The procedure is as follows: 

Содержание TC 1250-0.1 F

Страница 1: ...n E Mail info sauter eu Tel 49 0 7433 9933 199 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version 1 2 08 2017 GB Illustration TC 1250 0 1FN TC BA e 1712 PROFESSIONAL MEASURING ...

Страница 2: ...ge If you have any queries wishes or helpful suggestions do not hesitate to call our service number Models available TC 1250 0 1 F TC 1250 0 1 N TC 1250 0 1 FN TC_car 1250 0 1FN separate instruction manual available Summarize 1 Features 3 2 Specifications 3 3 Front Panel description 5 4 Measuring procedure 5 5 Zero Adjustment Calibration 6 6 Battery replacement 7 7 Adjustment foils 7 8 General adv...

Страница 3: ...d on magnetic metals e g steel iron nickel etc The N sensor measures the thickness of non magnetic coatings on non magnetic metals It is used on anodizing varnish paint enamel plastic coatings powder etc It can be applied on non magnetic materials like aluminium brass stainless steel etc The N sensor measures the thickness of non magnetic coatings on non magnetic metals It is used on anodization v...

Страница 4: ...V AAA UM 4 battery Operating conditions Temperature 0 to 50 C Humidity less than 80 Dimensions 127 x 67 x 28mm 5 0 x 2 6 x 1 1Inch Weight about 81g batteries not included Scope of delivery Carrying case Instruction manual According to types 1 Incorporated sensor at F model TC 1250 0 1 F with button F N 2 Incorporated sensor at N model TC 1250 0 1 N with button F N 3 Incorporated sensor at FN model...

Страница 5: ...ymbol Fe F for magnetic metals or NFe N for non magnetic metals which is indicated on the display It switches to the automatic measurement mode which also correctly identifies the zeroing plate or any other base material b The sensor 3 1 has to be placed onto a coating layer to be measured The reading on the display is the thickness of the coating layer This can be corrected by pressing the Plus k...

Страница 6: ...s the Power key 3 6 5 Zero Adjustment Calibration a Zero adjustment for Fe F and NFe should be carried out separately The Zero plate of iron has to be chosen if Fe is displayed and zero plate of alumi num has to be chosen if NFe is displayed Sensor 3 1 has to be placed carefully onto the appropriate base plate or another un coated base material The Zero key 3 3 has to be pressed without lifting th...

Страница 7: ...ise a The instrument should always be calibrated on the uncoated base material to be measured instead of the base plate included in the delivery Then accuracy is more precise right from the start b Sensors will eventually wear off Life of the sensor will depend on the number of measurements taken and how abrasive the coating is The change of a sensor should only be performed by qualified persons 9...

Страница 8: ...y be adjusted by professional persons Generally said The bigger the value of Ln the smaller the reading on the display for the same coat ing thickness Only a small change of the value of LN causes a big change in the reading of the upper measurement range at 500µm 20 mil The value of LN has to be adjusted as follows By pressing the Power key It lasts about 11 seconds from starting to press this ke...

Отзывы: