FH-M-BA-e-1718
7
BACK LIGHT:
PEAK: (Peak value)
Three functions available:
- Track mode (continuous measurement)
- Peak mode (capturing max. values)
- Auto-Peak mode, same as Peak-mode, but without the „Minimum limit value“
function
MEMORY: (Memory function)
Memorizes the Peak values to calculate the average value of the measurement
results (see chapter “memorizing Peak values”)
DELETE Function of memorized values (only active in „Memory“ mode)
PRINT: (Print function)
Sends the stored peak values to a PC or Printer (please see chapter 6.1)
High / Low Limit function
LEDs to display OK / NOT OK Tests
▼
Lower than lower limit
●
Lights if the STOP value is reached
▲
Higher than highest limit
A lower and an upper limit value can be defined. The instrument compares the
individual measuring results with the preset limit values and shows the OK or NOT
OK result by green or red light diode and by an acoustic sound.
To set these limit values, please see the SET Menu in section „Operation keys“.
Simple Measurement (Track Mode)
Displays (1) the actual operating force and force direction (7) (arrow)
To zero the display, press:
Peak-Hold Funktion (Peak Mode)
To change, please press: