9  Adjustment Procedure FH 

 

1. Switch on the 
instrument 

Press ON/OFF button 

The green light will be on 

2. Enter the 
adjustment procedure 

Directly after pressing ON/OFF, 
press PEAK and PRINT 
together for more times and  

very short intervals

 until the 

red light is on.  

The lower red light will be 
on 

3. Instrument type 

Press SET immediately, as 
soon as the red light turns off 
 

The max. Newton value of 
the instrument will be 
displayed, can be 
adjusted now respectively  

3a) 

(Back to the 

normal  mode ???)

 

(If you are in the operation 
mode again, the instrument has 
to be shut off. Start again with 
1. Eventually pressing the 
buttons faster) 

 

4. Select instrument 
type 

Please select the max. charge 
(N) of the instrument with the 

buttons▼▲  

The corresponding value 
to the instrument  appears 
in the display 

5. Save the settings 

Press SET  

 

6.   Enter the 
adjustment sequence 
 

Press MEMORY  

The red light on the right 
side lights up 

7.  Choose the 
adjustment weight 

Press UNIT and with▼▲ 
specify the calibration weight in 
Newton (X kg * 9,81) 
 

The weight is written in 
the display 

8.  Save the 
adjustment weight 
 

Press SET and UNIT 
simultaneously 

 

9.  Attach the weight 
to the instrument 

Attach the weight to the  
instrument and hold steady, 
then press ZERO  
 

The instrument changes 
to operation mode. 
Adjustment procedure is 
finished 

 

Содержание FH-M 1 kN

Страница 1: ...2336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Force Gauge SAUTER FH M V 1 8 06 2017 GB FH M BA e 1718 PROFESSIONAL MEASURING Sensor Outside ...

Страница 2: ...easuring instrument with its big functional range If you have any queries wishes or helpful suggestions do not hesitate to call our service number Summarize 1 Introduction 3 2 Scope of delivery 3 3 Working conditions 4 4 Specifications 4 5 Electrical Power Supply 4 6 Operation 5 6 1 Operation keys 5 7 Configuration of RS 232 interface 8 7 1 Output Protocol 9 8 Warnings 10 9 Adjustment Procedure FH...

Страница 3: ...nsor outside means that the load cell is outside the housing connected by a cable with the display unit SAUTER offers suitable software and a wide range of accessories optionally So you are able to configure your measuring instrument more versatile Please ask us at SAUTER or your SAUTER reseller or just visit us on our website www sauter eu 2 Scope of delivery SAUTER FH M incl rechargeable battery...

Страница 4: ...the instrument after an operating error You will find the description of mounting all force gauges on SAUTER Test stands in the instruction manuals of the correspondent test stands 3 Working conditions 10 C up to 30 C 15 up to 80 humidity 4 Specifications Measurement uncertainty 0 5 of Max measurement range Measurement frequency 2 000 Hz Weight 640 g display unit without external load cell 5 Elect...

Страница 5: ...attery 5 PEAK shows that Peak Hold mode is activated AUTO PEAK displays the Peak value only for a definite time 6 Average value respectively single Peak value 7 Indication of force direction 8 Occupancy of memory spaces 9 AVERAGE respectively memory mode 6 1 Operation keys ON OFF ON OFF key switch ON press about 1 sec UNIT Measurement units Press shortly select units between N kg and lb Press for ...

Страница 6: ...tand when reaching a limit force Adjustment of the force limit To change press or 1 x Press Auto OFF Function Turns the instrument off after a here defined period of time in sec P OFF To change press or Only active in battery Mode 1 x Press Peak Freeze Time PE 2E A PE HoldT period of time in sec in which a peak value is being shown in the display PEAK To change press or 1 x Press Optional data tra...

Страница 7: ... values to a PC or Printer please see chapter 6 1 High Low Limit function LEDs to display OK NOT OK Tests Lower than lower limit Lights if the STOP value is reached Higher than highest limit A lower and an upper limit value can be defined The instrument compares the individual measuring results with the preset limit values and shows the OK or NOT OK result by green or red light diode and by an aco...

Страница 8: ...ion keys Storage of Peak limit values and Average value calculation out of up to 10 measurement values Saving Peak values in the instrument Activation of AUTO PEAK Function by PEAK button Deactivation of Average Function by MEMORY button Now all peak values are stored automatically in the instrument s memory To browse through the stored values please use the or keys The values will be shown in the...

Страница 9: ...e PC by the ASCII Sign 9 The measured value that comes from the instrument has this format e g 0011 70 means 11 70 Newton if Newton is the selected unit ____ the other 6 places define the measured value as ASCII Signs _______ the first place describes the direction of the force 0 minus Pressure 1 plus Tension or 1021 15 means 21 15 N in tension direction _______ the first place describes the direc...

Страница 10: ...ple liquid slow draining off from a container suspended at the instrument Do not leave hanging a continuous load Overload Be sure to prevent overloading the instrument beyond the stated maximum load max minus any tare weight that may possibly exist This could damage the instrument risk of breakage Attention Always make sure that there are no people or materials below the load that could be injured...

Страница 11: ...quality assurance For this purpose the answerable user must define a suitable interval as well as the nature and scope of this check Information is available on the home page www KERN sohn com with regard to the monitoring of instrument test substances and the test weights required for this Test weights and instruments can be adjusted quickly and at a reasonable price in KERN s accredited DAkkS ca...

Страница 12: ...e instrument has to be shut off Start again with 1 Eventually pressing the buttons faster 4 Select instrument type Please select the max charge N of the instrument with the buttons The corresponding value to the instrument appears in the display 5 Save the settings Press SET 6 Enter the adjustment sequence Press MEMORY The red light on the right side lights up 7 Choose the adjustment weight Press ...

Отзывы: