FH-M-BA-e-1718
9
Pin Signal
Illustration
2
TxD
Output signal
3
RxD
Input signal
5
GND
Ground
6
+1.6 to +
2 V
>upper limit value
7
+1.6 to +
2 V
< lower limit value
8
+1.6 to +
2 V
OK
7.1 Output Protocol
RS-232 Parameter
-
Baudrate: 9600
-
Data-Bit: 8
-
Parity: none
-
Stop-bit: 1
The measured value is requested by the PC by the ASCII Sign “9”.
The measured value that comes from the instrument has this format:
e.g. 0011.70 means -11.70 Newton, if Newton is the selected unit
||-----|
| |____> the other 6 places define the measured value as ASCII-
Signs
|_______> the first place describes the direction of the force (0 =
minus = Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (in tension direction)
|_______> the first place describes the direction of the force (0 = minus =
Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (in tension direction)