15 

 

  When replacing the battery, remove the old battery (6F22, 006P). 

  Install a new

 

9V battery (6F22, 006P) by referring to the polarity marks (+ and -) on the battery 

compartment. 

  Place the battery cover on the unit and secure it to the unit with the screw. 

 
Note: When "BAT" is lit on the LCD, immediately replace all the battery with new ones. 

Otherwise the measurement accuracy may be affected or malfunction may occur. 

 
 

2. Installing a Vinyl Cover on the Unit 

To protect the SK-640PH main unit from water splashes, we recommend putting the vinyl cover on 
the unit before use. The cover also prevents the unit from becoming dirty. 
 

  Open the cover mouth and place the unit inside. When the unit is in the cover, its back side should 

be aligned with the Velcro closure side on the vinyl cover. 

  Pass the pH sensor connecting jack (BNC connecter) through the mouth of the vinyl cover, and 

then close the Velcro closure. 

  When removing the vinyl cover, disconnect the sensor probe first, and then remove in the reverse 

order of installation. 

 
Note: * The vinyl cover is not waterproof (splash-proof) according to the Japanese Industrial Standard   

* When the unit is contained in the vinyl cover, static electricity could build up on the vinyl cover

 

 

and cause pH readings to become unstable. In such cases, remove the cover before use. 

 
 

3. Connection of probe 

Before use, please connect the sensor probe following to the instruction manual of the probe.   

 

Connect probe to the main unit 

 

Connect the pH connecting terminal of sensor probe to the sensor connecting jack on top of   

        the main unit t.    Check that the pins of the jack are matched the groove of pH pH connecting 

terminal of sensor probe 
 

Connect the   

 
 
 
 
 
 
 
 
 

4. Auto power-off function 

The unit is turned off by the auto power-off function if no key is operated for 20 minutes. Turn off the 
auto power-off beforehand if measurement is to be done continuously.   
 
* Release the Auto power-off function 

 

Press the PWR key while pressing the CAL key at the same time to turn on the unit, and then 

   

  release the PWR key.   

Содержание SK-640PH

Страница 1: ...Instruction manual for No 6422 00 Handy type pH meter Model SK 640PH No 3 Nishi Fukudacho Kanda Chiyodaku Tokyo 101 0037 Japan URL http www sksato co jp english N 11...

Страница 2: ...e If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may increase Never disassemble or modify the unit Doing so may cau...

Страница 3: ...mperature compensation Features Large LCD for easy reading The pH value and the temperature are indicated on a large LCD Auto Power Off The auto power off function turns off the power automatically if...

Страница 4: ...e and MIN value The maximum and minimum voltage mV can also be logged To return to the measurement mode press the REC key for three seconds or more while in the MAX MIN logging mode Pressing this key...

Страница 5: ...minimum pH or voltage value is displayed BAT Lights up when battery power is insufficient ATC Lights up when the temperature sensor is connected Display of time voltage Indicates the time elapsed sin...

Страница 6: ...g the vinyl cover disconnect the sensor probe first and then remove in the reverse order of installation Note The vinyl cover is not waterproof splash proof according to the Japanese Industrial Standa...

Страница 7: ...on the right 2 Installing the glass electrode protection bottle 2 1 Install the protection bottle to the sensor probe in the reverse order of removal 2 2 Hold the sensor probe upright so as not to spi...

Страница 8: ...fter replacing the battery c When very accurate measurement is necessary For more accurate measurement calibration of about once a day is recommended For calibrating the SK 620PH prepare the following...

Страница 9: ...alibration value is 4 04 7 Press the REC key to store the calibration value in the unit memory SA is displayed on the LCD to indicate that the calibration data has been stored in the unit memory Now t...

Страница 10: ...ssible to perform calibration using the oxalate standard solution pH 1 68 at 25 C or borate standard solution pH 9 18 at 25 C CAL key REC key REC Key REC Key Measurement mode Calibration pH4 Calibrati...

Страница 11: ...hich is capable of measuring both pH and temperature is used connect both the BNC plug pH and the diode plug temperature 2 Press the PWR key to turn on the unit The whole LCD lights up for approximate...

Страница 12: ...21...

Страница 13: ...eft unwashed the liquid measured tends to form a film on the surface of the electrode causing measurement errors Always wash the pH electrode every time after use Washing can be done with water distil...

Страница 14: ...bout the RS 232C output RS 232C output This instrument outputs data in real time on the pH voltage and temperature while in the measurement mode or the MAX MIN logging mode When the unit is connected...

Страница 15: ...d replacing battery The power to the unit is turned off The auto power off is turned on Cancel the auto power off on the main unit Before measuring Error code pH display section The sensor probe is no...

Страница 16: ...an electric conductivity of approx 100 S cm or more The temperature of the sample largely fluctuated Make the sample temperature stable The pH values change with temperature Readings are abnormal The...

Страница 17: ...unication conditions or protocols are not correct Check the communication conditions or protocols About the RS 232C output The serial port on the computer is not open Check that the serial port is ope...

Страница 18: ...ent Temperature 0 to 50 Humidity lower than 80 rh no condensing Battery Life About 20 hours continuous Materials Case ABS resin Electrode Glass Probe Epoxy resin Dimensions Main unit W 72 H 182 D 31 5...

Страница 19: ...28 6400 40 PH40 Carrying case 6400 50 PH50 Electrode protection bottle...

Отзывы: