20 

 

Relation between pH value and temperature 

 

When performing calibration, set the pH value according to the temperature of the standard solution 
as follows; 

Standard solution 

Name of solution 

Temp. 

pH value 

Temp. 

pH value 

pH 4 

 

Phthalate solution 

 

4.00 

30

 

4.02 

 

4.00 

35

 

4.02 

10

 

4.00 

40

 

4.04 

15

 

4.00 

45

 

4.05 

20

 

4.00 

50

 

4.06 

25

 

4.01 

pH 7 

Neutral

 

phosphate solution   

 

6.98 

30

 

6.85 

 

6.95 

35

 

6.84 

10

 

6.92 

40

 

6.84 

15

 

6.90 

45

 

6.83 

20

 

6.88 

50

 

6.83 

25

 

6.86 

pH 10 

Carbonate solution 

 

10.32 

30

 

9.97 

 

10.24 

35

 

9.92 

10

 

10.18 

40

 

9.89 

15

 

10.12 

45

 

9.86 

20

 

10.06 

50

 

9.83 

25

 

10.01 

                                                                                                                                            JIS Z 8802-2011 

 

Measuring procedure 

 

(1) Connect the sensor probe to the main unit. When the PHP-31 sensor, which is capable of 

measuring both pH and temperature, is used, connect both the BNC plug (pH) and the diode plug 
(temperature). 

 
(2) Press the PWR key to turn on the unit. The whole LCD lights up for approximately three seconds. 

The unit enters the measurement mode. 

 
(3) Remove the glass electrode protection bottle from the sensor probe. 
 
(4) Immerse the sensor probe at least 3 cm deep into the material to be measured. 
 
(5) Read the pH value when the reading becomes stable. 
 
(6) To display the voltage, press the pH/mV key. The voltage detected by the sensor probe is 

displayed. 

 
(7) To end the measurement, press the PWR key to turn off the unit. 

Wash the glass electrode with water (distilled, pure or tap water), install the glass electrode 
protection bottle and store the instrument. 

 
Note: For how to install or remove the glass electrode protection bottle, refer to "Glass electrode 

protection bottle". 

Содержание SK-640PH

Страница 1: ...Instruction manual for No 6422 00 Handy type pH meter Model SK 640PH No 3 Nishi Fukudacho Kanda Chiyodaku Tokyo 101 0037 Japan URL http www sksato co jp english N 11...

Страница 2: ...e If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may increase Never disassemble or modify the unit Doing so may cau...

Страница 3: ...mperature compensation Features Large LCD for easy reading The pH value and the temperature are indicated on a large LCD Auto Power Off The auto power off function turns off the power automatically if...

Страница 4: ...e and MIN value The maximum and minimum voltage mV can also be logged To return to the measurement mode press the REC key for three seconds or more while in the MAX MIN logging mode Pressing this key...

Страница 5: ...minimum pH or voltage value is displayed BAT Lights up when battery power is insufficient ATC Lights up when the temperature sensor is connected Display of time voltage Indicates the time elapsed sin...

Страница 6: ...g the vinyl cover disconnect the sensor probe first and then remove in the reverse order of installation Note The vinyl cover is not waterproof splash proof according to the Japanese Industrial Standa...

Страница 7: ...on the right 2 Installing the glass electrode protection bottle 2 1 Install the protection bottle to the sensor probe in the reverse order of removal 2 2 Hold the sensor probe upright so as not to spi...

Страница 8: ...fter replacing the battery c When very accurate measurement is necessary For more accurate measurement calibration of about once a day is recommended For calibrating the SK 620PH prepare the following...

Страница 9: ...alibration value is 4 04 7 Press the REC key to store the calibration value in the unit memory SA is displayed on the LCD to indicate that the calibration data has been stored in the unit memory Now t...

Страница 10: ...ssible to perform calibration using the oxalate standard solution pH 1 68 at 25 C or borate standard solution pH 9 18 at 25 C CAL key REC key REC Key REC Key Measurement mode Calibration pH4 Calibrati...

Страница 11: ...hich is capable of measuring both pH and temperature is used connect both the BNC plug pH and the diode plug temperature 2 Press the PWR key to turn on the unit The whole LCD lights up for approximate...

Страница 12: ...21...

Страница 13: ...eft unwashed the liquid measured tends to form a film on the surface of the electrode causing measurement errors Always wash the pH electrode every time after use Washing can be done with water distil...

Страница 14: ...bout the RS 232C output RS 232C output This instrument outputs data in real time on the pH voltage and temperature while in the measurement mode or the MAX MIN logging mode When the unit is connected...

Страница 15: ...d replacing battery The power to the unit is turned off The auto power off is turned on Cancel the auto power off on the main unit Before measuring Error code pH display section The sensor probe is no...

Страница 16: ...an electric conductivity of approx 100 S cm or more The temperature of the sample largely fluctuated Make the sample temperature stable The pH values change with temperature Readings are abnormal The...

Страница 17: ...unication conditions or protocols are not correct Check the communication conditions or protocols About the RS 232C output The serial port on the computer is not open Check that the serial port is ope...

Страница 18: ...ent Temperature 0 to 50 Humidity lower than 80 rh no condensing Battery Life About 20 hours continuous Materials Case ABS resin Electrode Glass Probe Epoxy resin Dimensions Main unit W 72 H 182 D 31 5...

Страница 19: ...28 6400 40 PH40 Carrying case 6400 50 PH50 Electrode protection bottle...

Отзывы: