Introduction 

 
Thank you for purchasing the handy type pH meter Model SK-640PH.   

This  product  is  designed  to  measure  hydrogen  ion  concentration  (pH)  and  temperature  (liquid 
temperature). Do not use this instrument for any other purposes.   

Read  this  instruction  manual  before  using  the  SK-640PH  and  keep  it  in  a  safe  place  for  future 
references. 

 
 

Important safety instructions 

 

 

The SK-640PH is not explosion-proof. Never use it for flammable materials. 

 

DANGER 

RISK OF EXPLOSION 

 

Do not use the pH standard solutions for any purposes other than to calibrate pH meters. 
The solutions are not for drinking. Never try to drink it. If accidentally ingested, consult a doctor 
immediately.   

For  any  other  inquiries,  contact  the  store  where  you  purchased  the  instrument  or  our  service 
network. 

 
 
           

Caution 

 

For proper use of the SK-640PH, be sure to observe the followings: 
- This is a precision instrument. Be careful not to drop or knock it. 
- The operating temperature range for this instrument is 0 to 50°C and a relative humidity of less than 

80%. Operating this product outside the temperature range may result in unit malfunctions. 

- Avoid using this unit in a place exposed to direct sunlight or near a heat source. Otherwise, the case 

may become deformed or malfunction may be caused. 

- If this instrument is left in a car in the hot summer sun, it will become extremely hot and may 

malfunction. Do not leave the unit in such a place. 

- If this unit is used in an environment where electrical noise is generated, the display may become 

unstable or the measurement error may increase. 

- Never disassemble or modify the unit. Doing so may cause malfunctions. 
- Do not use this instrument outside the operating range. Doing so may cause malfunctions.   
- This instrument is not waterproof. Do not place it in water or use it where it may be splashed with water. 
- When used outdoors, keep the unit out of rain or water splashes. If it becomes wet, the unit may fail. 
- If the unit will not be used for a long period of time, always remove the battery. Otherwise, the 

battery power may be wasted and the battery fluid may leak, resulting in malfunctions. 

- Keep the battery out of reach of children. If a child swallows a battery accidentally, consult a doctor 

immediately. 

- Do not dispose of used battery in a fire. 
- For environmental protection, dispose of used battery in compliance with local rules and regulations. 
- Do not wash or wipe this unit with alcohol, thinner, or other solvents. If the unit becomes dirty, wipe 

it with a tightly-wrung towel or the like that has been dipped in warm water with a neutral detergent. 

- Do not alter, pull forcibly, bend or bind the sensor probe, as doing so may cause a break in the 

probe. Also, do not place heavy objects on the cord or heat it, as doing so damage the cord. 

- When measuring, do not let the sensor probe come in contact with the human body. 
- When the unit is used for the first time or after it has been stored for several days, the pH 

measurement response may be slower. In such cases, immerse the sensor probe electrode in 

Содержание SK-640PH

Страница 1: ...Instruction manual for No 6422 00 Handy type pH meter Model SK 640PH No 3 Nishi Fukudacho Kanda Chiyodaku Tokyo 101 0037 Japan URL http www sksato co jp english N 11...

Страница 2: ...e If this unit is used in an environment where electrical noise is generated the display may become unstable or the measurement error may increase Never disassemble or modify the unit Doing so may cau...

Страница 3: ...mperature compensation Features Large LCD for easy reading The pH value and the temperature are indicated on a large LCD Auto Power Off The auto power off function turns off the power automatically if...

Страница 4: ...e and MIN value The maximum and minimum voltage mV can also be logged To return to the measurement mode press the REC key for three seconds or more while in the MAX MIN logging mode Pressing this key...

Страница 5: ...minimum pH or voltage value is displayed BAT Lights up when battery power is insufficient ATC Lights up when the temperature sensor is connected Display of time voltage Indicates the time elapsed sin...

Страница 6: ...g the vinyl cover disconnect the sensor probe first and then remove in the reverse order of installation Note The vinyl cover is not waterproof splash proof according to the Japanese Industrial Standa...

Страница 7: ...on the right 2 Installing the glass electrode protection bottle 2 1 Install the protection bottle to the sensor probe in the reverse order of removal 2 2 Hold the sensor probe upright so as not to spi...

Страница 8: ...fter replacing the battery c When very accurate measurement is necessary For more accurate measurement calibration of about once a day is recommended For calibrating the SK 620PH prepare the following...

Страница 9: ...alibration value is 4 04 7 Press the REC key to store the calibration value in the unit memory SA is displayed on the LCD to indicate that the calibration data has been stored in the unit memory Now t...

Страница 10: ...ssible to perform calibration using the oxalate standard solution pH 1 68 at 25 C or borate standard solution pH 9 18 at 25 C CAL key REC key REC Key REC Key Measurement mode Calibration pH4 Calibrati...

Страница 11: ...hich is capable of measuring both pH and temperature is used connect both the BNC plug pH and the diode plug temperature 2 Press the PWR key to turn on the unit The whole LCD lights up for approximate...

Страница 12: ...21...

Страница 13: ...eft unwashed the liquid measured tends to form a film on the surface of the electrode causing measurement errors Always wash the pH electrode every time after use Washing can be done with water distil...

Страница 14: ...bout the RS 232C output RS 232C output This instrument outputs data in real time on the pH voltage and temperature while in the measurement mode or the MAX MIN logging mode When the unit is connected...

Страница 15: ...d replacing battery The power to the unit is turned off The auto power off is turned on Cancel the auto power off on the main unit Before measuring Error code pH display section The sensor probe is no...

Страница 16: ...an electric conductivity of approx 100 S cm or more The temperature of the sample largely fluctuated Make the sample temperature stable The pH values change with temperature Readings are abnormal The...

Страница 17: ...unication conditions or protocols are not correct Check the communication conditions or protocols About the RS 232C output The serial port on the computer is not open Check that the serial port is ope...

Страница 18: ...ent Temperature 0 to 50 Humidity lower than 80 rh no condensing Battery Life About 20 hours continuous Materials Case ABS resin Electrode Glass Probe Epoxy resin Dimensions Main unit W 72 H 182 D 31 5...

Страница 19: ...28 6400 40 PH40 Carrying case 6400 50 PH50 Electrode protection bottle...

Отзывы: